FDA Adverse Event Malfunction Summary report: N

COMBUR-10-M 100 STR

MDR report key: 8711688 · Received June 19, 2019

Report

Report Number
1823260-2019-02243
Event Type
Malfunction
Date Received
June 19, 2019
Date of Event
May 29, 2019
Report Date
September 10, 2019
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JMT
PMA / PMN Number
K032437
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXPIRATION DATE WAS UPDATED. THE RETENTION MATERIAL, LOT 37378000, WAS MEASURED ON INVESTIGATION UNIT URISYS 1800 AND U 411 ANALYZER WITH 0-NATIVE URINE AND A NITRITE-DILUTION-SERIES AND WAS VISUALLY CHECKED. THE RESULTS OF THE MEASUREMENTS FULFILL THE REQUIREMENTS. NO FALSE POSITIVE RESULTS WERE OBSERVED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

AFTER CALIBRATION, THERE WERE NO MORE DISCREPANCIES WITH NITRITE RESULTS. THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem · 1

THE INITIAL REPORTER STATED THAT THEY HAVE RECEIVED DISCREPANT HIGH RESULTS FOR AN UNSPECIFIED NUMBER PATIENT SAMPLES TESTED FOR NITRITES ON A COBAS U 411 ANALYZER. NO INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER STATED THAT THEY TESTED 142 URINE SAMPLES PER DAY OVER A 6 DAY PERIOD USING THE SAME TEST STRIP LOT NUMBER. WHEN TESTING THE SAMPLES ON THE U 411 ANALYZER, THE NITRITE RESULTS WILL BE POSITIVE. WHEN RUNNING TEST STRIPS MANUALLY ON THESE SAMPLES, THE VISUAL READING IS NEGATIVE FOR NITRITE. NO ADVERSE EVENTS WERE ALLEGED TO HAVE OCCURRED WITH THE PATIENTS. THE SERIAL NUMBER OF THE U 411 ANALYZER IS (B)(4). THE REPORTER'S PRODUCT WAS REQUESTED FOR INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
505344 COMBUR-10-M 100 STR URINE TEST STRIPS JMT ROCHE DIAGNOSTICS COMBUR-10-M 37378004

Patients

Seq Age Sex Outcome Treatment
1