PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH
Report
- Report Number
- 2032227-2019-18358
- Event Type
- Malfunction
- Date Received
- June 19, 2019
- Date of Event
- June 12, 2019
- Report Date
- July 1, 2020
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169507227
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE RECEIVED WITH CRACKED DISPLAY WINDOW CORNERS AND CRACKED BATTERY TUBE THREADS NOTED. THE TEST RESERVOIR P-CAP WAS ABLE TO LOCK IN PLACE. DEVICE PASSED DISPLACEMENT TEST. HOWEVER DEVICE ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. UNABLE TO VERIFY POSSIBLE UNDER DELIVERY ANOMALY OR PERFORM PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT ACCURACY TEST, A21 ERROR TEST AND OCCLUSION TEST DUE TO MOTOR ERROR. DEVICE PASSED SELF TEST, OFF NO POWER TEST AND ALL OPERATING CURRENTS TESTED WITHIN THE SPECIFICATION. DEVICE WAS DOWNLOADED TO CARELINK PRO PROPERLY.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP WAS UNDER DELIVERING. THE CUSTOMER'S BLOOD GLUCOSE WAS 240, 220, 250 MG/DL. THE INSULIN PUMP WAS CRACK ON THE BACK OF THE INSULIN PUMP. THE TROUBLESHOOTING WAS PERFORMED FOR THE DAMAGE. THE PRODUCT WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 505328 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAH | B5723NAHJ | 00643169507227 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |