FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH

MDR report key: 8711343 · Received June 19, 2019

Report

Report Number
2032227-2019-18358
Event Type
Malfunction
Date Received
June 19, 2019
Date of Event
June 12, 2019
Report Date
July 1, 2020
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169507227
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE RECEIVED WITH CRACKED DISPLAY WINDOW CORNERS AND CRACKED BATTERY TUBE THREADS NOTED. THE TEST RESERVOIR P-CAP WAS ABLE TO LOCK IN PLACE. DEVICE PASSED DISPLACEMENT TEST. HOWEVER DEVICE ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. UNABLE TO VERIFY POSSIBLE UNDER DELIVERY ANOMALY OR PERFORM PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT ACCURACY TEST, A21 ERROR TEST AND OCCLUSION TEST DUE TO MOTOR ERROR. DEVICE PASSED SELF TEST, OFF NO POWER TEST AND ALL OPERATING CURRENTS TESTED WITHIN THE SPECIFICATION. DEVICE WAS DOWNLOADED TO CARELINK PRO PROPERLY.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP WAS UNDER DELIVERING. THE CUSTOMER'S BLOOD GLUCOSE WAS 240, 220, 250 MG/DL. THE INSULIN PUMP WAS CRACK ON THE BACK OF THE INSULIN PUMP. THE TROUBLESHOOTING WAS PERFORMED FOR THE DAMAGE. THE PRODUCT WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
505328 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH B5723NAHJ 00643169507227

Patients

Seq Age Sex Outcome Treatment
1 52 YR