FDA Adverse Event Malfunction Summary report: N

CLEARCUT INTREPID KNIVES

MDR report key: 8706942 · Received June 18, 2019

Report

Report Number
2523835-2019-00261
Event Type
Malfunction
Date Received
June 18, 2019
Report Date
September 19, 2019
Manufacturer
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
Product Code
HNN
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

NO KNIFE SAMPLE HAS BEEN RETURNED FOR EVALUATION FOR THE REPORT OF BLADES WERE DULL THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. A REVIEW OF THE DEVICE HISTORY RECORD RELATED TO THE REPORTED LOT NUMBER INDICATES THAT THE PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE PRODUCT¿S ACCEPTANCE CRITERIA. A COMPLAINT HISTORY EXAMINATION FOR DEVICE COMPONENT LOTS TRACEABLE TO THE REPORTED LOT NUMBER INDICATES THAT THERE ARE NO ADDITIONAL COMPLAINTS ASSOCIATED WITH THE COMPONENT LOTS FOR THE REPORTED ISSUE. A PHOTO ATTACHED TO THE PARENT COMPLAINT WAS REVIEWED BY THE MANUFACTURING SITE. THE PHOTO CONFIRMS THE REPORTED CUSTOM PAK NUMBER AND CUSTOM PAK LOT NUMBER FOR THIS QS FILE AND QS (B)(4). A SAMPLE WAS NOT RETURNED AND THE DEVICE HISTORY RECORD REVIEW OF THE LOT NUMBER PROVIDED INDICATES THAT THE PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO ACCEPTANCE CRITERIA THEREFORE, THE ROOT CAUSE FOR THE CUSTOMER COMPLAINT ISSUE CANNOT BE DETERMINED. THE EXACT ROOT CAUSE FOR THIS COMPLAINT IS UNKNOWN THEREFORE, SPECIFIC ACTION WITH REGARDS TO THIS COMPLAINT CANNOT BE TAKEN. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. COMPLAINTS ARE REVIEWED AND MONITORED AT REGULAR INTERVALS FOR ANY SIGNIFICANT ADVERSE TRENDS. NO ADDITIONAL ACTION IS REQUIRED AT THIS TIME. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 1

NO SAMPLE HAS BEEN RECEIVED BY MANUFACTURING FOR EVALUATION. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ADDITIONAL INFORMATION HAS BEEN REQUESTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 1

A HEALTHCARE PROFESSIONAL REPORTED THAT MULTIPLE KNIFE BLADES WERE DULL DURING SURGERIES. PATIENT IMPACT INFORMATION IS UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED CONFIRMED THAT THE DULL KNIVES WERE NOTED DURING SEPARATE UNCONFIRMED SURGERY DATES. THE DULL KNIVES REQUIRED ADDITIONAL PUSH IN ORDER TO CUT. THERE WAS NO IMPACT TO ANY PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
503871 CLEARCUT INTREPID KNIVES KNIFE, OPHTHALMIC HNN ALCON RESEARCH, LLC - ALCON PRECISION DEVICE NA 255408M

Patients

Seq Age Sex Outcome Treatment
1