FDA Adverse Event Malfunction Summary report: N

O-ARM O2 IMAGING SYSTEM

MDR report key: 8705210 · Received June 17, 2019

Report

Report Number
3004785967-2019-01077
Event Type
Malfunction
Date Received
June 17, 2019
Date of Event
May 22, 2019
Report Date
July 31, 2019
Manufacturer
MEDTRONIC NAVIGATION, INC (LITTLETON)
Product Code
OWB
PMA / PMN Number
K151000
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D11: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: BI71000444, SERIAL/LOT #: REV. 1 : S/N (B)(4) , UBD: , UDI#: H2) DEVICE EVALUATION: H3, H6) THE ROTOR CONTROLLER BOX WAS RETURNED TO MEDTRONIC FOR EVALUATION. THE REPORTED COMPLAINT WAS CONFIRMED THROUGH FUNCTIONAL TESTING, PERFORMANCE TESTING, AND VISUAL/PHYSICAL EXAMINATION. THERE WAS AN ELECTRICAL FAILURE WITH THE ROTOR CONTROLLER BOX. IT WAS DETERMINED THAT THE ROTOR CONTROLLER BOX WAS FAULTY. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

A MEDTRONIC REPRESENTATIVE (REP) TESTED AND SERVICED THE EQUIPMENT AT THE SITE. MECHANICAL TESTING FAILED DUE TO THE X-DRIVE, Y-DRIVE, AND Z-DRIVE NOT WORKING. THE ROTOR CONTROLLER BOX WAS REPLACED AND THE FIRMWARE WAS UPDATED, THUS RESOLVING THE FAILURE. THE IMAGING SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION REGARDING AN IMAGING DEVICE. IT WAS REPORTED THAT DURING AN UPDATE TO THE SYSTEM, INITIALIZATION OF THE MOTION FAILED DUE TO A DEFECTIVE ROTOR CONTROLLER. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498475 O-ARM O2 IMAGING SYSTEM INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB MEDTRONIC NAVIGATION, INC (LITTLETON) BI70002000

Patients

Seq Age Sex Outcome Treatment
1 "SEE H10...."