FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 8704205 · Received June 17, 2019

Report

Report Number
9617229-2019-06680
Event Type
Injury
Date Received
June 17, 2019
Date of Event
March 25, 2019
Report Date
June 17, 2019
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628030169
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INFORMATION CONTAINED IN THIS REPORT WAS PREVIOUSLY SUBMITTED THROUGH PSR ON 22APR2019. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. DEVICE EVALUATION, VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: WEIGHT TO THE SPEC, DEFORMATION. CLOUDY AND BUBBLES WERE OBSERVED AFTER AUTOCLAVE DISINFECTION PROCESS. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING PROCESS. NO OPENINGS OR ISSUES RELATED TO RUPTURE DEVICE WERE OBSERVED. REASON FOR REOPERATION WAS DUE TO RUPTURE. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498416 INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3119109 10888628030169

Patients

Seq Age Sex Outcome Treatment
1 34 YR Required Intervention