SENSOR ENLITE MMT-7008A
Report
- Report Number
- 2032227-2019-16864
- Event Type
- Malfunction
- Date Received
- June 17, 2019
- Date of Event
- June 8, 2019
- Report Date
- December 18, 2019
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OZO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
INSPECTED ONE OPENED AND USED SENSOR AND PERFORMED CONTINUITY RESISTANCE TEST AND SENSOR FAILED TEST. FOUND CANNULA BENT UNABLE TO CONFIRM CUSTOMER RECEIVED SENSOR IN SAID CONDITION DUE TO CUSTOMER RETURNED OPENED AND USED.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 140 MG/DL AND THE SENSOR GLUCOSE WAS 65 MG/DL AT THE TIME OF THE INCIDENT. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 496491 | SENSOR ENLITE MMT-7008A | ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND | OZO | MEDTRONIC MINIMED | MMT-7008A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |