FDA Adverse Event Malfunction Summary report: N

SENSOR ENLITE MMT-7008A

MDR report key: 8702900 · Received June 17, 2019

Report

Report Number
2032227-2019-16864
Event Type
Malfunction
Date Received
June 17, 2019
Date of Event
June 8, 2019
Report Date
December 18, 2019
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED SENSOR AND PERFORMED CONTINUITY RESISTANCE TEST AND SENSOR FAILED TEST. FOUND CANNULA BENT UNABLE TO CONFIRM CUSTOMER RECEIVED SENSOR IN SAID CONDITION DUE TO CUSTOMER RETURNED OPENED AND USED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 140 MG/DL AND THE SENSOR GLUCOSE WAS 65 MG/DL AT THE TIME OF THE INCIDENT. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
496491 SENSOR ENLITE MMT-7008A ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC MINIMED MMT-7008A

Patients

Seq Age Sex Outcome Treatment
1 36 YR