FDA Adverse Event Injury Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 8702870 · Received June 17, 2019

Report

Report Number
2032227-2019-16853
Event Type
Injury
Date Received
June 17, 2019
Date of Event
May 1, 2019
Report Date
August 25, 2019
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
00763000090203
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE PASSED THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME OR SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST, AND THE DELIVERY ACCURACY TEST AT 0.0864 INCHES. NO BATTERY ISSUES NOTED DURING TESTING. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED LOW BLOOD GLUCOSE LEVEL. CUSTOMER'S BLOOD GLUCOSE VALUE WAS 38 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DECLINED FOR LOW BLOOD GLUCOSE. THE CUSTOMER WAS TREATED WITH JUICE. THE DEVICE WILL BE RETURNED FOR THE ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
496232 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG2N0B0 00763000090203

Patients

Seq Age Sex Outcome Treatment
1 49 YR Other FRN-UNK-RSVR, UNOMED INF SET