ABBOTT COMMANDER FPC
Report
- Report Number
- 1628664-2007-00309
- Event Type
- Other
- Date Received
- June 25, 2007
- Date of Event
- January 23, 2007
- Report Date
- June 4, 2007
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- JQW
- PMA / PMN Number
- K894069
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION OF SERVICE TEXT FROM COMPLAINT. THE FIELD SERVICE ENGINEER PERFORMED ROUTINE CLEANING, CALIBRATING, AND ADJUSTMENT PROCEDURES FOR THE BARCODE READER. A REVIEW WAS PERFORMED OF THE COMPLAINT TEXT AND SERVICE HISTORY FOR THE COMMANDER FPC. FIELD SERVICE PERFORMED ROUTINE MAINTENANCE PROCEDURES FOR CLEANING, CALIBRATING AND ADJUSTING THE BARCODE READER. THE ABBOTT COMMANDER FLEXIBLE PIPETTING CENTER OPERATIONS MANUAL, SECTION 10, TROUBLESHOOTING AND DIAGNOSTICS, TROUBLESHOOTING, A BARCODE HAS BEEN MISREAD SECTION: CONTAINS INFORMATION ABOUT TROUBLESHOOTING A MISREAD BARCODE. THIS IS A FINAL REPORT.
THE ACCOUNT STATED THAT UPON REVIEW OF COMMANDER FPC RECORDS OBSERVED BARCODE MISREADS WITH COMMANDER FPC WHEN PIPETTING HIV-1/2 EIA, HTLV-I/II EIA AND HCV 2.0 EIA ASSAYS. THERE WERE A TOTAL OF 8 BARCODE MISREADS BETWEEN TWO COMMANDER FPC SERIAL NUMBERS BETWEEN 2006 AND 2007. THIS REPORT IS FOR BARCODE MISREAD #2 WHERE BARCODE ID 8561219 READ AS 6561219 IN 2007. THE SAME BARCODE ID WAS ALSO PROCESSED ON THE PRISM ANALYZER WITH CORRECT BARCODE IDENTIFICATION. NO SPECIFIC PATIENT INFORMATION IS AVAILABLE. THE ACCOUNT RECOGNIZED THE INCORRECT BARCODE ID PRIOR TO RESULTS BEING REPORTED. NO INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. SERVICE WAS REQUESTED FOR THE COMMANDER FPC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ABBOTT COMMANDER FPC | AUTOMATED SYSTEM FOR THE PROCESSING OF EIAS | JQW | ABBOTT MANUFACTURING, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK YR | FPC LIST 3A46-47| HTLV-I/II| HIV-1/2 EIA| HCV 2.0 EIA |