FDA Adverse Event Malfunction Summary report: N

SYR 3ML 22GA 1-1/4IN JPK/SIL NO ASSY NDL

MDR report key: 8699858 · Received June 14, 2019

Report

Report Number
9614033-2019-00165
Event Type
Malfunction
Date Received
June 14, 2019
Date of Event
May 30, 2019
Report Date
June 21, 2019
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: PHOTOS RECEIVED FOR INVESTIGATION. UPON EVALUATION, IT IS OBSERVED THE PRESENCE OF A HAIR IN THE BLISTERPACK. THE LOT WAS INSPECTED ACCORDING TO THE CURRENT WORK INSTRUCTION WITH SATISFACTORY RESULTS FOR THE CHARACTERISTICS REQUIRED FOR ITS APPROVAL INCLUDING FUNCTIONAL TESTS. NO NON-CONFORMING PART THAT COULD BE ATTRIBUTABLE TO THE DEFECT REPORTED BY THE CLIENT WAS PRESENTED. NO DEVIATIONS (QUALITY NOTIFICATIONS) WERE REPORTED DURING THE MANUFACTURING PROCESS. AS A PART OF THE PERSONAL PROTECTION EQUIPMENT, COVER HEAD IS USED, WHICH COVERS THE HEAD COMPLETELY. PRIOR TO THIS THE HAIR INCIDENT WAS DUE TO A FAIL IN THE CLOTHING PROCEDURE BY THE OPERATIONAL STAFF.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SYR 3ML 22GA 1-1/4IN JPK/SIL NO ASSY NDL PACKAGING SEAL WAS FOUND WITH A HAIR TRAPPED IN IT. LOT #'S 9015772 AND 9015795 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH TO ENGLISH: "SYRINGE WITH HAIR TRAPPED IN THE PACKING SEAL."

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9015772. MEDICAL DEVICE EXPIRATION DATE: 2024-01-12. DEVICE MANUFACTURE DATE: 2019-01-15. MEDICAL DEVICE LOT #: 9015795. MEDICAL DEVICE EXPIRATION DATE: 2024-01-31. DEVICE MANUFACTURE DATE: 2019-01-15. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYR 3 ML 22GA 1-1/4 IN JPK/SIL NO ASSY NDL PACKAGING SEAL WAS FOUND WITH A HAIR TRAPPED IN IT. LOT #'S 9015772 AND 9015795 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH TO ENGLISH: "SYRINGE WITH HAIR TRAPPED IN THE PACKING SEAL."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
495621 SYR 3ML 22GA 1-1/4IN JPK/SIL NO ASSY NDL SYRINGE FMF BECTON DICKINSON DE MEXICO SEE SECTION H.10.

Patients

Seq Age Sex Outcome Treatment
1 Other