IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM
Report
- Report Number
- 0002023141-2019-00272
- Event Type
- Injury
- Date Received
- June 14, 2019
- Date of Event
- April 9, 2019
- Report Date
- July 8, 2019
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- PMA / PMN Number
- K061410
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- DENTIST
Narratives
SIMILAR COMPLAINTS FOR THE INABILITY TO ACHIEVE PRIMARY STABILITY (UNABLE TO PLACE IMPLANT INTO OSTEOTOMY) HAVE BEEN PREVIOUSLY INVESTIGATED. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORDS REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH HAVE NOT SUGGESTED THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. TO DATE, ALL COMPLAINT DATA WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. A REVIEW OF THE SUBJECT LOT (1220397) DID NOT IDENTIFY ANY RELATED NON-CONFORMANCES WHICH WOULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. ADDITIONALLY, THERE HAVE BEEN NO RELATED COMPLAINTS REPORTED AGAINST THE SUBJECT LOT. THESE FACTS DO NOT SUGGEST A SYSTEMIC QUALITY ISSUE WITH THE LOT. REPORTED EVENTS LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. A ROOT CAUSE FOR THE COMPLAINT COULD NOT BE DETERMINED WITH THE INFORMATION PROVIDED AS THE EVENT CANNOT BE RECREATED. COMPLAINT IS THEREFORE NON-VERIFIABLE. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: D4: (B)(4).
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
ZIMMER BIOMET COMPLAINT (B)(4). PATIENT'S WEIGHT NOT PROVIDED/UNKNOWN. ADDITIONAL 510K NUMBERS: K011028 AND K013227.
IT WAS REPORTED THAT THE IMPLANT (TSVWB8) DID NOT ACHIEVE PRIMARY STABILITY IN THE BONE. NO OTHER IMPLANT WAS ABLE TO BE PLACED DURING THE SAME PROCEDURE. THE WOUND WAS CLOSED AND AWAITING HEALING FOR NEW IMPLANT PLACEMENT. TOOTH LOCATION 19.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 495493 | IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1220397 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Required Intervention |