FDA Adverse Event Malfunction Summary report: N

BD 1ML SYRINGE LUER-LOK TIP

MDR report key: 8693257 · Received June 12, 2019

Report

Report Number
1213809-2019-00614
Event Type
Malfunction
Date Received
June 12, 2019
Date of Event
May 24, 2019
Report Date
July 3, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: TWO LOOSE 1ML SYRINGES AND ONE OPENED AND EMPTY BLISTER PACK FROM BATCH #9042756 (P/N 309628) WERE RECEIVED AND EVALUATED. THE SYRINGES WERE DISASSEMBLED, AND NO VISUAL DEFECTS WERE OBSERVED. THE BD 1ML HYPODERMIC SYRINGE IS A PISTON SYRINGE UNDER PRODUCT CODE FMF, AS DESCRIBED AT 21 C.F.R. § 880.5860, INTENDED FOR USE BY HEALTH CARE PROFESSIONALS FOR GENERAL PURPOSE FLUID ASPIRATION AND INJECTION. BD HAS ONLY TESTED AND VALIDATED THIS PRODUCT FOR USE IN ACCORDANCE WITH THE REFERENCED PRODUCT CODE. THE REPORTED DEFECT WAS NOT IDENTIFIED IN THE SAMPLES RECEIVED. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER OCCURRED WITH A BD 1ML SYRINGE LUER-LOK¿ TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, . "WHEN INJECT WITH 1ML LL SYRINGE,MIXED HISTOACRYL(GLUE)+LIPIODOL ULTRA SOLUTION(CONTRAST MEDIA)+5%DEXTROSE, HISTOACRYL BECOMES HARD IN 1ML LL SYRINGE." 3 OCCURRENCES REPORTED BEFORE USE.

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOREIGN MATTER OCCURRED WITH A BD 1ML SYRINGE LUER-LOK¿ TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN INJECT WITH 1ML LL SYRINGE, MIXED HISTOACRYL (GLUE)+LIPIODOL ULTRA SOLUTION(CONTRAST MEDIA)+5%DEXTROSE, HISTOACRYL BECOMES HARD IN 1ML LL SYRINGE." 3 OCCURRENCES REPORTED BEFORE USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
487043 BD 1ML SYRINGE LUER-LOK TIP SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 9042756 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 Other