HEAL COLLAR 4.5X5.5, 3MM
Report
- Report Number
- 0002023141-2019-00259
- Event Type
- Malfunction
- Date Received
- June 12, 2019
- Date of Event
- April 29, 2019
- Report Date
- September 26, 2019
- Manufacturer
- ZIMMER DENTAL
- Product Code
- NHA
- PMA / PMN Number
- K111852
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
Narratives
ZIMMERBIOMET COMPLAINT NUMBER (B)(4). THE HEAL COLLAR 4.5X5.5, 3MM (ITEM # HC453) WAS RETURNED FOR INVESTIGATION. VISUAL INSPECTION OF THE RETURNED PRODUCT DID NOT IDENTIFIED ANY MALFUNCTION. FUNCTIONAL TESTING WAS PERFORMED FOR THE RETURNED PRODUCT THE HEALING COLLARS DID NOT DISPLAY A MALFUNCTION AND THREADED EFFECTIVELY WHEN TESTED WITH AN IN-HOUSE COMPATIBLE IMPLANT. APPROPRIATE DOCUMENTATION WAS REVIEWED AND THE FOLLOWING INFORMATION WAS IDENTIFIED: INSTRUCTIONS FOR USE FOR HEALING COLLARS, HEALING SCREWS, SURGICAL COVER SCREWS AND TEMPORARY GINGIVAL CUFFS 9658 REV1-01/15 INFORMATION IDENTIFIED: DESCRIPTION AND ADVERSE EFFECTS A DEVICE HISTORY REVIEW WAS PERFORMED AND NO RELATED NONCONFORMANCES WERE NOTED. ALSO, A COMPLAINT HISTORY SEARCH WAS PERFORMED USING OUR COMPLAINT HANDLING SYSTEM AND THERE WERE NO ADDITIONAL RELATED COMPLAINTS FOR THIS PRODUCT LOT. COMPLAINANT REPORTED TWO HEALING COLLARS (ADDITIONAL HEALING COLLAR WAS REPORTED ON (MFR - 0002023141 - 2019 - 00261- 1) WOULD NOT SCREW INTO THE IMPLANT. THE REPORTED EVENT COULD NOT BE CONFIRMED. A DEFINITIVE ROOT CAUSE FOR THIS COMPLAINT COULD NOT BE DETERMINED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT IDENTIFIER UNKNOWN / NOT PROVIDED. AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. PATIENT SEX UNKNOWN / NOT PROVIDED. WEIGHT UNKNOWN / NOT PROVIDED. LOT NUMBER UNKNOWN / NOT PROVIDED.
IT WAS REPORTED THAT THE HEALING COLLAR (HC453) WOULD NOT SEAT INSIDE THE IMPLANT. ANOTHER HEALING COLLAR WAS USED TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 484436 | HEAL COLLAR 4.5X5.5, 3MM | HEALING COLLAR | NHA | ZIMMER DENTAL | 2017091223 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |