FDA Adverse Event Malfunction Summary report: N

GENERAL MIS SPHERE INSTRUMENT SET

MDR report key: 8691478 · Received June 12, 2019

Report

Report Number
3005180920-2019-00454
Event Type
Malfunction
Date Received
June 12, 2019
Date of Event
May 13, 2019
Report Date
June 12, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
UDI-DI
07630030882890
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 06 JUN 2019: GENERAL MIS SPHERE INSTRUMENT SET REF 11.01002, LOT 1901032: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01-MAR-2019. EXPIRATION DATE: 2024-02-18. NO ANOMALIES FOUND RELATED TO THE ISSUE. TO DATE, (B)(4) ITEMS OF THIS LOT HAVE ALREADY BEEN SOLD, WITHOUT ANY OTHER SIMILAR EVENT REGISTERED. CONSIDERING THE SPECIFIC COMPONENT FEMORAL IMPACTOR/ EXTRACTOR, REF77.11.0063, LOT MAT27910 ((B)(4) PIECES): THIS IS THE 4TH CASE RECEIVED FOR THE BREAKAGE OF THE CENTRAL BODY. THE OTHER COMPLAINTS ARE: (B)(4) [MDR2019-00448], (B)(4) AND (B)(4). PRELIMINARY INVESTIGATION PERFORMED BY R&D PROJECT MANAGER: THE BREAKAGE OF A PIECE OF THE FEMORAL IMPACTOR EXTRACTOR HEAD COULD HAVE BEEN CAUSED DUE SEVERAL FACTORS SUCH AS AN EXCESSIVE TIGHTENING FORCE APPLIED DURING FEMORAL IMPLANT FIXATION, EXCESSIVE FORCE/IMPACTIONS TRYING TO ADJUST THE TRIAL FEMORAL POSITIONS (ML AND FLEXUM POSITIONING).

Description of Event or Problem · 1

DURING THE PRIMARY KNEE SURGERY AND WHILE IMPACTING THE FEMORAL COMPONENT, THE GMK EFFICIENCY FEMORAL IMPACTOR/EXTRACTOR BROKE AT THE CENTRAL PART. NO BROKEN FRAGMENTS FELL INTO THE PATIENT'S SURGICAL WOUND AND THERE WAS ONE MINUTE DELAY IN THE CASE. THE SURGEON SWAPPED TO CONVENTIONAL INSTRUMENTATION TO COMPLETE THE CASE AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
487428 GENERAL MIS SPHERE INSTRUMENT SET SURGICAL INSTRUMENT FOR KNEE SURGERY LXH MEDACTA INTERNATIONAL SA 1901032 07630030882890

Patients

Seq Age Sex Outcome Treatment
1 Other