FDA Adverse Event Malfunction Summary report: N

JAG PRECURSOR

MDR report key: 869115 · Received June 22, 2007

Report

Report Number
6000123-2007-00028
Event Type
Malfunction
Date Received
June 22, 2007
Date of Event
February 1, 2007
Report Date
February 27, 2007
Manufacturer
BOSTON SCIENTIFIC
Product Code
EZB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL EXAM OF THE RETURNED DEVICE REVEALED THAT APPROXIMATELY 10-15% OF THE PEBAX WAS DETACHED FROM THE CORE WIRE. SEVERAL LOCATIONS OF THE CORE WIRE WERE EXPOSED; THESE LOCATIONS SHOWED EVIDENCE OF COATING ADHESIVE. THE GUIDEWIRE ALSO HAD TWO KINKS APPROX 1.5 CM AND 4.0 CM FROM THE DISTAL END. THE PEBAX OUTER DIAMETER, IN UNDAMAGED SECTIONS, OF THE WIRE WAS CONFIRMED TO BE WITHIN SPEC. THE PHYSICAL EVIDENCE DISPLAYED BY THE DEVICE INDICATES THE DAMAGE WAS LIKELY CAUSED BY AN INTERACTION WITH ANOTHER DEVICE, IN CONJUNCTION WITH USER HANDLING. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED ON THE PERTINENT LOT; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE FOR SIMILAR COMPLAINTS WAS CONDUCTED; ONE ADDITIONAL COMPLAINT FROM THE SAME FACILITY, WITH THE SAME FAILURE MODE WAS RECORDED FOR THIS LOT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED ON THE PERTINENT LOT; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE FOR SIMILAR COMPLAINTS WAS CONDUCTED; ONE ADDITIONAL COMPLAINT FROM THE SAME FACILITY, WITH THE SAME FAILURE MODE, WAS RECORDED FOR THIS LOT. THE MAY 2007 15-MONTH JAGWIRE PRODUCT FAMILY COMPLAINT TREND REPORT, INCLUSIVE OF ALL FAILURE MODES, WAS REVIEWED; NO ADVERSE TREND WAS NOTED AND NO DATA POINT EXCEEDED ANY COMPLAINT ALERT LIMIT. A CAPA FOR TIP SEPARATION ISSUES IS IN PROGRESS.

Description of Event or Problem · 1

NOTE: THIS MEDWATCH REPORT IS BEING SUBMITTED AS A RESULT OF RETURNED DEVICE EVAL. (SEE H10). IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION IN 2007, THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY ON A FEMALE PT, THE PHYSICIAN ATTEMPTED TO "DELIVER THE GUIDEWIRE THROUGH THE SPHINCTEROTOME [WHEN] THE WIRE'S TIP COVER FELL APART SHOWING THE METAL INSIDE". THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN DEVICE. PT'S CONDITION WAS REPORTED AS "GOOD".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 JAG PRECURSOR EZB EZB BOSTON SCIENTIFIC M0055658011 8078181

Patients

Seq Age Sex Outcome Treatment
1 65 YR