JAG PRECURSOR
Report
- Report Number
- 6000123-2007-00028
- Event Type
- Malfunction
- Date Received
- June 22, 2007
- Date of Event
- February 1, 2007
- Report Date
- February 27, 2007
- Manufacturer
- BOSTON SCIENTIFIC
- Product Code
- EZB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
Narratives
VISUAL EXAM OF THE RETURNED DEVICE REVEALED THAT APPROXIMATELY 10-15% OF THE PEBAX WAS DETACHED FROM THE CORE WIRE. SEVERAL LOCATIONS OF THE CORE WIRE WERE EXPOSED; THESE LOCATIONS SHOWED EVIDENCE OF COATING ADHESIVE. THE GUIDEWIRE ALSO HAD TWO KINKS APPROX 1.5 CM AND 4.0 CM FROM THE DISTAL END. THE PEBAX OUTER DIAMETER, IN UNDAMAGED SECTIONS, OF THE WIRE WAS CONFIRMED TO BE WITHIN SPEC. THE PHYSICAL EVIDENCE DISPLAYED BY THE DEVICE INDICATES THE DAMAGE WAS LIKELY CAUSED BY AN INTERACTION WITH ANOTHER DEVICE, IN CONJUNCTION WITH USER HANDLING. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED ON THE PERTINENT LOT; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE FOR SIMILAR COMPLAINTS WAS CONDUCTED; ONE ADDITIONAL COMPLAINT FROM THE SAME FACILITY, WITH THE SAME FAILURE MODE WAS RECORDED FOR THIS LOT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED ON THE PERTINENT LOT; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE FOR SIMILAR COMPLAINTS WAS CONDUCTED; ONE ADDITIONAL COMPLAINT FROM THE SAME FACILITY, WITH THE SAME FAILURE MODE, WAS RECORDED FOR THIS LOT. THE MAY 2007 15-MONTH JAGWIRE PRODUCT FAMILY COMPLAINT TREND REPORT, INCLUSIVE OF ALL FAILURE MODES, WAS REVIEWED; NO ADVERSE TREND WAS NOTED AND NO DATA POINT EXCEEDED ANY COMPLAINT ALERT LIMIT. A CAPA FOR TIP SEPARATION ISSUES IS IN PROGRESS.
NOTE: THIS MEDWATCH REPORT IS BEING SUBMITTED AS A RESULT OF RETURNED DEVICE EVAL. (SEE H10). IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION IN 2007, THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY ON A FEMALE PT, THE PHYSICIAN ATTEMPTED TO "DELIVER THE GUIDEWIRE THROUGH THE SPHINCTEROTOME [WHEN] THE WIRE'S TIP COVER FELL APART SHOWING THE METAL INSIDE". THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN DEVICE. PT'S CONDITION WAS REPORTED AS "GOOD".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | JAG PRECURSOR | EZB | EZB | BOSTON SCIENTIFIC | M0055658011 | 8078181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |