ARCHITECT STAT TROPONIN-I
Report
- Report Number
- 1415939-2007-00172
- Event Type
- Other
- Date Received
- June 19, 2007
- Date of Event
- June 12, 2007
- Report Date
- June 13, 2007
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- MMI
- PMA / PMN Number
- K041192
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NT
- Reporter Occupation
- UNKNOWN
Narratives
AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATES, THAT ONE PT SAMPLE GENERATED AN INITIAL ARCHITECT STAT TROPONIN-I ASSAY RESULT OF 1.20 UG/L. SIX HRS LATER, ANOTHER SAMPLE WAS DRAWN AND GENERATED A STAT TROPONIN-I ASSAY RESULT OF LESS THAN 0.01 UG/L. THE FOLLOWING DAY, ANOTHER SAMPLE WAS DRAWN AND GENERATED A STAT TROPONIN-I ASSAY RESULT OF 0.02 UG/L. THE CUSTOMER THEN REPEATED THE SAMPLES THAT GENERATED INITIAL RESULTS OF 1.20 UG/L AND LESS THAN 0.01 UG/L AND RECEIVED RESULTS OF 0.02 UG/L AND 0.00 UG/L RESPECTIVELY. CONTROLS HAVE BEEN WITHIN SPECIFICATIONS ON ALL RUNS. THE PT UNDERWENT A SURGICAL PROCEDURE BASED ON THE INITIAL ELEVATED RESULT OF 1.20 UG/L. THERE IS NO FURTHER PT INFO AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT STAT TROPONIN-I | CMIA FOR MEASUREMENT OF CARDIAC TROPONIN-I | MMI | ABBOTT LABORATORIES | NA | 13037UN06 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK YR | Other | ARCHITECT I2000SR ANALYZER LIST# 3M74-01 |