FDA Adverse Event Malfunction Summary report: N

BD SAFETYGLIDE¿ NEEDLE

MDR report key: 8683745 · Received June 10, 2019

Report

Report Number
1213809-2019-00602
Event Type
Malfunction
Date Received
June 10, 2019
Date of Event
May 18, 2019
Report Date
May 23, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMI
UDI-DI
30382903059028
PMA / PMN Number
K951254
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: SEVEN SEALED SAFETYGLIDE NEEDLES WERE RECEIVED, CONFIRMED TO BE FROM BATCH #8143729 (MATERIAL 305902). EACH NEEDLE WAS TAKEN OUT OF THE PACKAGE AND VISUALLY EVALUATED. NO VISUAL DEFECTS WERE OBSERVED. EACH NEEDLE WAS THEN ATTACHED TO A 3ML SYRINGE AND HAD ITS SAFETY SHIELD ACTIVATED TO TEST FUNCTION. ALL SAFETY SHIELDS ACTIVATED AS EXPECTED WITHOUT ISSUES. NO DEFECTS WERE OBSERVED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. INVESTIGATION CONCLUSION: THE REPORTED DEFECT WAS NOT IDENTIFIED IN THE SAMPLES RECEIVED. ROOT CAUSE DESCRIPTION: N/A. RATIONALE: THE REPORTED DEFECT WAS NOT IDENTIFIED IN THE SAMPLES RECEIVED.

Description of Event or Problem · 1

MATERIAL NO.: 305902, BATCH NO.: UNKNOWN. IT WAS REPORTED THAT AFTER USE OF THE BD SAFETYGLIDE¿ NEEDLE THE SAFETY MECHANISM WAS STUCK AND RESISTING TO LOCK RESULTING IN A NEEDLE STICK TO THE LEFT THUMB OF THE NURSE. NURSE RECEIVED BLOOD WORK AND IS TAKING PROPHYLAXIS; PATIENT ALSO AGREED TO BLOOD WORK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: SAFETY MECHANISM WAS STUCK AND RESISTING TO LOCK; NEEDLE STICK OCCURRED ON THE LEFT THUMB WITH A USED NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
478870 BD SAFETYGLIDE¿ NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON MEDICAL SYSTEMS 8143729 30382903059028

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention