FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 8682491 · Received June 10, 2019

Report

Report Number
8010042-2019-00369
Event Type
Malfunction
Date Received
June 10, 2019
Report Date
October 29, 2019
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K180098
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED TANDEM USE OF A HIGH FREQUENCY JET VENTILATOR (HFJV) IMPLIES THAT THE HFJV IS CONNECTED TO AN ET TUBE ADAPTER AND IT INDUCES A ¿JET¿ OF GAS OUT OF THE ADAPTER INTO THE AIRWAY. THE MEASURED INSPIRATORY VALUES WOULD BE AS PER THE SET PARAMETER VALUES BUT THE MEASURED EXPIRATORY VALUES WOULD INCLUDE THE VALUES OF THE HFJV AND THEREFORE, THE MEASURED EXPIRATORY VALUES WOULD NOT REFLECT THE PARAMETER SETTINGS AND WILL LEAD TO GENERATION OF APPROPRIATE ALARMS, FOR EXAMPLE HIGH O2 CONCENTRATION. THE USE OF A HFJV IN TANDEM WITH THE VENTILATOR HAS NOT BEEN TESTED, APPROVED, OR RECOMMENDED FOR USE WITH THIS VENTILATOR SYSTEM. THEREFORE THE CONSEQUENCES OF ITS USE WITH OUR VENTILATOR ARE UNKNOWN. AS NO LOG FILES FROM EVENT DATE WERE PROVIDED AND NO PARTS WERE RETURNED FOR OUR INVESTIGATION, IT HAS NOT BEEN POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE REPORTED EVENT BUT THE USE OF HFJV IN TANDEM WITH THE VENTILATOR IS THE MOST LIKELY CONTRIBUTING FACTOR. PROBLEM CODE: 4117.

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION: E2018003. GETINGE USA SALES, LLC, 45 BARBOUR POND DRIVE, WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Description of Event or Problem · 0

MANUFACTURER'S REF # : (B)(4).

Description of Event or Problem · 0

A USER FACILITY MEDWATCH REFERENCE # (B)(4) WAS RECEIVED STATING THAT: "NICU INFANT (PREEMIE) ON JET AND SERVO. ON SERVO VENTILATOR WHEN THE FIO2 WAS ON 45% AND THE MEASURED FIO2 READINGS WAS 50%. THERE WAS AN ALARM FOR HIGH O2 CONCENTRATION."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476532 SERVO-U VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1 0 YR