ALTRX +4 10D 40IDX58OD
Report
- Report Number
- 1818910-2019-95555
- Event Type
- Injury
- Date Received
- June 10, 2019
- Date of Event
- May 15, 2019
- Report Date
- May 21, 2019
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- LPH
- UDI-DI
- 10603295016571
- PMA / PMN Number
- K062148
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER RELATED REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT CODE/LOT NUMBER COMBINATION SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS, THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. PER INTERNAL PROCEDURES, THE EVENT INFORMATION AND ANY INVESTIGATIONAL INPUTS RECEIVED AS PART OF REQUIRED FOLLOW UP WERE REVIEWED. FOR THIS INVESTIGATION, NO IMMEDIATE ACTION WAS REQUIRED AS NO ALLEGED DEFICIENCY WITH THE DEVICE WAS IDENTIFIED. MONITOR ¿ EXEMPT PER WI-7915 - KNOWN POSSIBLE COMPLICATION OF JOINT REPLACEMENT SURGERY. NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. IN ORDER TO DETERMINE IF A LOT RELATED ISSUE WAS POSSIBLE, A WORLDWIDE COMPLAINT DATABASE SEARCH WAS PERFORMED. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND ONE ADDITIONAL REPORT (COM-133646) AGAINST THE PROVIDED ALTRX +4 10D 40IDX58OD PRODUCT CODE 122140158, LOT NUMBER 473438 COMBINATION. THE ADDITIONAL REPORT WAS FOR A PATIENT BEING REVISED TO ADDRESS INFECTION, AND IT IS NOT CONSIDERED TO BE RELATED TO THE CURRENT REPORTED EVENT OF A PATIENT EXPERIENCING PAIN AND DISLOCATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO ADDITIONAL RELATED REPORTS AGAINST THE PROVIDED PRODUCT CODE/LOT NUMBER COMBINATION. BASED ON THE INABILITY TO FIND ANY ADDITIONAL RELATED REPORTS AGAINST THE PROVIDED PRODUCT CODE/LOT NUMBER COMBINATION, IT IS REASONABLE TO CONCLUDE THAT THERE ARE NO ANOMALIES WITH REGARD TO MANUFACTURING OR INSPECTION CONTAINED IN THE DEVICE HISTORY RECORDS THAT WOULD CONTRIBUTE TO THE REPORTED EVENT. VISUAL EXAMINATION OF THE PROVIDED X-RAY IMAGES COULD NOT CONFIRM THE REPORTED EVENT. THE FEMORAL HEAD WAS CENTRALLY LOCATED WITHIN THE ACETABULAR COMPONENTS. NO EVIDENCE OF IMPLANT FRACTURE, DISASSOCIATION, OR ANYTHING INDICATIVE OF A DEVICE NON-CONFORMANCE WAS FOUND. THE REPORTED EVENT IS CONSIDERED ONE OF THE POSSIBLE COMPLICATIONS OF JOINT REPLACEMENT. INVESTIGATIONAL INPUTS WERE REQUESTED AS INDICATED PER INTERNAL PROCEDURES FOR THIS FAILURE MODE. THE COMPLAINT INFORMATION PROVIDED HAS BEEN REVIEWED FOR COMPLAINT CODING, MEDICAL DEVICE REPORTING, AND OTHER DATA REQUIRED BY THE COMPLAINT SYSTEM. FOLLOW-UP FOR ADDITIONAL EVENT INFORMATION, IF APPLICABLE, WAS CONDUCTED UTILIZING WORK INSTRUCTION WI-7915 APPENDIX A. - WITHOUT THE PHYSICAL COMPLAINT SAMPLE ASSOCIATED WITH THIS REPORT, IT WAS NOT POSSIBLE TO DETERMINE IF THE DEVICE FAILED TO MEET SPECIFICATIONS AT THE TIME IT WAS RELEASED FOR DISTRIBUTION. - THE DEVICE ASSOCIATED WITH THIS EVENT WAS USED IN THE TREATMENT OF THE PATIENT AS PRESCRIBED BY THE PRESIDING SURGEON. - FROM THE EVENT INFORMATION RECEIVED, IT WAS NOT POSSIBLE TO DETERMINE THE RELATIONSHIP OF THE DEVICE TO THE REPORTED EVENT. NO EVIDENCE WAS FOUND INDICATING PRODUCT ERROR WAS A CONTRIBUTING FACTOR. THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFO BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT NULL DEVICE HISTORY BATCH NULL DEVICE HISTORY REVIEW NULL DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. CORRECTED: D6 (IMPLANTATION DATE) ADDED: (EXPLANTATION DATE), (NO CODE AVAILABLE FOR SURGICAL INTERVENTION) IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT #: (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER RELATED REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT CODE/LOT NUMBER COMBINATION SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS, THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM.
PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE SECOND DISLOCATION WAS CORRECTED BY CLOSED REDUCTION IN THE EMERGENCY DEPARTMENT. EACH TIME THE PATIENT DISLOCATED AFTER BENDING OVER. THE PHYSICIAN REPORTED THE PATIENT WAS EXPERIENCING PAIN IN THE RIGHT HIP. THERE IS NO EVIDENCE OF A SECOND REVISION AFTER THE FIRST REVISION ON (B)(6) 2014, CAPTURED ON (B)(4).
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY ==> NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER RELATED REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT CODE/LOT NUMBER COMBINATION SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS, THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
CLINICAL ADVERSE EVENT RECEIVED FOR RECURRENT DISLOCATIONS EVENT IS SERIOUS AND IS CONSIDERED MODERATE EVENT IS DEFINITELY RELATED TO DEVICE AND THERE IS A REMOTE POSSIBILITY THAT IT IS RELATED TO PROCEDURE. DOI: (B)(6) 2012; DOR: (B)(6) 2014 (INSTABILITY/DISLOCATION--HEAD, STEM, AND LINER REVISED); DOE: (B)(6) 2019; (RIGHT HIP).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 476316 | ALTRX +4 10D 40IDX58OD | PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS | LPH | DEPUY ORTHOPAEDICS INC US | 1221-40-158 | 473438 | 10603295016571 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |