FDA Adverse Event Malfunction Summary report: N

MICROSCAN PROMPT INOCULATION SYSTEM-D

MDR report key: 868113 · Received September 29, 2006

Report

Report Number
2919016-2006-00023
Event Type
Malfunction
Date Received
September 29, 2006
Date of Event
September 5, 2006
Report Date
September 7, 2006
Manufacturer
DADE BEHRING, INC.
Product Code
JSD
Removal / Correction Number
FCA 000264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EVALUATION CODES: PERFORMANCE TESTING OF CUSTOMER RETURNS AND RETENTION SAMPLES. RESULTS: IN-HOUSE TESTING PERFORMED AND CONFIRMED INSUFFICIENT GROWTH ISSUE WITH CUSTOMER PROVIDED SAMPLES AND ALSO WITH RETENTION SAMPLES. CONCLUSION: OTHER - DADE BEHRING HAS INITIATED A FIELD CORRECTION FOR THIS ISSUE AND NOTIFIED CUSTOMERS OF THE POTENTIAL FOR DISCREPANT MIC AND/OR IDENTIFICATION RESULTS WITH CLINICAL AND QC ISOLATES WITH THE IDENTIFIED PROMPT INOCULATION SYSTEM-D LOT.

Description of Event or Problem · 1

ACCOUNT REPORTED TO DADE BEHRING THAT THEY RETESTED A CLINICAL ISOLATE THAT THEY PREVIOUSLY OBSERVED INSUFFICIENT GROWTH WITH. CLINICAL ISOLATE YIELDED SUSCEPTIBLE ERYTHROMYCIN AND CLINDAMYCIN RESULTS WHEN USING THE IDENTIFIED PROMPT LOT. ALTERNATE PROMPT LOT GAVE RESISTANT RESULTS FOR ERYTHROMYCIN AND CLINDAMYCIN. ACCOUNT DID NOT REPORT THE SUSCEPTIBLE RESULTS TO THE DR. NO REPORT OF INJURY OR ILLNESS ASSOCIATED WITH THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROSCAN PROMPT INOCULATION SYSTEM-D PANEL INOCULATING SUPPLIES JSD DADE BEHRING, INC. NA 20071127

Patients

Seq Age Sex Outcome Treatment
1 YR