MICROSCAN PROMPT INOCULATION SYSTEM-D
Report
- Report Number
- 2919016-2006-00024
- Event Type
- Malfunction
- Date Received
- September 29, 2006
- Date of Event
- September 15, 2006
- Report Date
- September 15, 2006
- Manufacturer
- DADE BEHRING, INC.
- Product Code
- JSD
- Removal / Correction Number
- FCA 000264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EVALUATION: PERFORMANCE TESTING OF CUSTOMER RETURNS AND RETENTION SAMPLES. RESULTS: IN-HOUSE TESTING PERFORMED AND CONFIRMED INSUFFICIENT GROWTH ISSUE WITH CUSTOMER PROVIDED SAMPLES AND ALSO WITH RETENTION SAMPLES. CONCLUSION: DADE BEHRING HAS INITIATED A FIELD CORRECTION FOR THIS ISSUE AND NOTIFIED CUSTOMERS OF THE POTENTIAL FOR DISCREPANT MIC AND/OR IDENTIFICATION RESULTS WITH CLINICAL AND QC ISOLATES WITH THE IDENTIFIED PROMPT INOCULATION SYSTEM-D LOT.
ACCOUNT REPORTED TO DADE BEHRING THAT THEY HAVE OBSERVED INTERMITTENT INSUFFICIENT GROWTH (NO RESULTS). HAVE OBSERVED NO GROWTH IN THE GROWTH WELL OF MICROSCAN PANELS (NO RESULTS). REPORT WAS ASSOCIATED WITH IDENTIFIED PROMPT LOT AND RESOLVED WITH ALTERNATE LOT OF PROMPT. NO REPORT OF INJURY OR ILLNESS ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSCAN PROMPT INOCULATION SYSTEM-D | PANEL INOCULATING SUPPLIES | JSD | DADE BEHRING, INC. | NA | 20071127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |