STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2019-05190
- Event Type
- Injury
- Date Received
- June 6, 2019
- Date of Event
- May 7, 2019
- Report Date
- July 26, 2019
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: FLAT CREASES. A LEAK TEST AND MICROSCOPIC ANALYSIS WERE PERFORMED WHICH IDENTIFIED: A >1 MM VOID IN NON-TEXTURED, VALVE-TO SHELL-BOND AREA. THE FILL TEST INSPECTION WAS PERFORMED AND THE RESULT WAS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO LEAK WAS OBSERVED IN THE DEVICE.
HEALTHCARE PROFESSIONAL REPORTED A RIGHT-SIDE CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. THE REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE UNKNOWN. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.
HEALTHCARE PROFESSIONAL REPORTED A RIGHT-SIDE CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 468986 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 1735536 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Required Intervention |