SYRINGE 1ML S/T W/NDL 25X5/8 RB
Report
- Report Number
- 1213809-2019-00592
- Event Type
- Malfunction
- Date Received
- June 5, 2019
- Date of Event
- May 17, 2019
- Report Date
- July 10, 2019
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 30382903096269
- PMA / PMN Number
- K980580
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: ONE PHOTO AND THREE 1ML SYRINGES WITH NEEDLES IN FULLY SEALED BLISTER PACKS CONFIRMED TO BE FROM BATCH #9053640 (P/N 309626) WERE RECEIVED AND EVALUATED. IT WAS OBSERVED ALL THE SYRINGES HAD A JAMMED STOPPER CONDITION AND WERE REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE JAMMED STOPPER DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. IT WAS MOST LIKELY CAUSED BY A MISALIGNMENT BETWEEN THE PLUNGER AND BARREL DIALS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
IT WAS REPORTED THAT TWO SYRINGE 1ML S/T W/NDL 25X5/8 RB EXPERIENCED A MISALIGNED/JAMMED STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 309626 BATCH NO.: 9053640 VERBATIM: THE SYRINGES WERE FOUND WITH THE PLUNGER MISPLACED.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT TWO SYRINGE 1ML S/T W/NDL 25X5/8 RB EXPERIENCED A MISALIGNED/JAMMED STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 309626, BATCH NO.: 9053640. VERBATIM: THE SYRINGES WERE FOUND WITH THE PLUNGER MISPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 463539 | SYRINGE 1ML S/T W/NDL 25X5/8 RB | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 9053640 | 30382903096269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |