DRIVER SHAFT, T-15, LONG
Report
- Report Number
- 1220246-2019-01132
- Event Type
- Injury
- Date Received
- June 5, 2019
- Date of Event
- May 8, 2019
- Report Date
- June 5, 2019
- Manufacturer
- ARTHREX, INC.
- Product Code
- LXH
- UDI-DI
- 00888867225176
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- 501
Narratives
COMPLAINT CONFIRMED, THE DISTAL END WAS FOUND TO BE TWISTED AND TO HAVE A .060 PIECE BROKEN FROM THE DEVICE. IT IS STATED A TORQUE INDICATING ADAPTER WAS USED WHICH IS RECOMMENDED TO PREVENT OVER-TORQUING. AS THE DEVICE WAS FOUND TO MEET DIMENSIONAL AND MATERIAL SPECIFICATIONS, THE MOST LIKELY CAUSE(S) OF THIS TYPE OF EVENT INCLUDE CONTINUALLY APPLYING TORQUE WHEN THE IMPLANT IS NOT PROPERLY ALIGNED, LEVERAGING, OR TORQUING WHILE LEVERAGING THE DEVICE AND/OR IMPROPER USE OF THE TORQUE INDICATING ADAPTER.
IT WAS REPORTED THAT THE FOLLOWING OCCURRED DURING A REVISION PROCEDURE TO REMOVE A COMPETITOR¿S DEVICE (ARTHROSURFACE HEMI CAP). AFTER REMOVAL OF THE COMPETITOR¿S DEVICE A REVERSE TOTAL SHOULDER PROCEDURE WAS THEN PERFORMED. DURING THE PROCEDURE AN ARTHREX AR-1999HH (RATCHETING HANDLE), LOT 8001640 WAS NOT RATCHETING CORRECTLY AND COULD NOT BE USED. WHILE BEING USED TO INSERT A CENTRAL SCREW, AN AR-9545-T15-03 (DRIVER SHAFT), LOT 8001643, TIP BROKE OFF IN THE HEAD OF THE SCREW. THE TIP WAS NOT RETRIEVED AND IT REMAINS IN THE SCREW IN THE PATIENT. AN AR-9545-T15-02, (DRIVER SHAFT), LOT 8001819, WAS ALSO USED AND IT WAS REPORTED THAT THE THREADS OF THE DRIVER WERE TWISTED WHEN TIGHTENING THE SUPERIOR SCREW. THE SUPERIOR PERIPHERAL SCREW AR-9145-42NL, LOT UNKNOWN, SNAPPED OFF AT THE SHAFT AND HALF OF THE SCREW WAS LEFT IN THE PATIENT UNRETRIEVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 463871 | DRIVER SHAFT, T-15, LONG | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | ARTHREX, INC. | DRIVER SHAFT, T-15, LONG | 8001643 | 00888867225176 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |