FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 8670741 · Received June 5, 2019

Report

Report Number
2245578-2019-00139
Event Type
Malfunction
Date Received
June 5, 2019
Date of Event
May 19, 2019
Report Date
July 3, 2019
Manufacturer
ABBOTT POINT OF CARE
Product Code
MMI
UDI-DI
10054749001986
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT #: (B)(4). THE INVESTIGATION WAS COMPLETED ON 07/02/2019. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED AND RETURNED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AD (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CTNI CARTRIDGE LOT K19052

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4).APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON (B)(6) 2019, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT TROPONIN RESULT OF 1.02 ON A (B)(6) YEAR OLD FEMALE PATIENT WITH CHEST PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. (B)(6). AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THERE WAS NO FINAL DIAGNOSIS PROVIDED. THE CUSTOMER STATES THE PATIENT WAS NOT SENT TO THE CATH LAB, NOR WAS THERE EKG RESULTS AVAIL. THERE IS NO EVIDENCE THAT THE PATIENT SUFFERED AN MI. THE INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL: ART: 714258-00Q, REPORTABLE RANGE: 0.00 - 50.00, REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS), REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESULTS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464576 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE NA K19052 10054749001986

Patients

Seq Age Sex Outcome Treatment
1 39 YR