FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER

MDR report key: 8668595 · Received June 4, 2019

Report

Report Number
1024879-2019-00960
Event Type
Malfunction
Date Received
June 4, 2019
Date of Event
May 14, 2019
Report Date
August 2, 2019
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686563
PMA / PMN Number
K011984
Removal / Correction Number
PAS-19-1567
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: INVESTIGATION SUMMARY: AS OF (B)(6)2019 THIS COMPLAINT IS UNDER THE SCOPE OF FIELD ACTION # PAS-19-1567: BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR THE SAFETY MECHANISM NOT RETRACTING WITH THE INCIDENT LOT WAS NOT OBSERVED. TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND NO ISSUES RELATED TO RETRACTION WERE OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH CAPA#982194. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR THE SAFETY MECHANISM NOT RETRACTING WITH THE INCIDENT LOT WAS NOT OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH CAPA#982194. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. ROOT CAUSE DESCRIPTION: CAPA#982194 HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. RATIONALE: BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. H.6. ADDITIONAL INFORMATION: THIS INFORMATION IS ASSOCIATED WITH FIELD ACTION # 2243072-07/22/2019-009-R 1. RECALL SUMMARY BD IS CONDUCTING A VOLUNTARY MEDICAL DEVICE RECALL FOR CERTAIN LOTS OF BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER, CATALOG# 368656. BD HAS CONFIRMED THAT THESE LOTS MAY EXHIBIT SEPARATION OF FRONT AND REAR BARRELS UPON ACTIVATION OF THE SAFETY FEATURE THAT RETRACTS THE NEEDLE. 2. PRODUCT AND SCOPE BD IS CONDUCTING A VOLUNTARY MEDICAL DEVICE RECALL FOR THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER, CATALOG NUMBER 368356, FOR THE FOLLOWING 12 LOTS: 9079767 9079770 9081829 9086811 9086812 9091607 9091626 9094659 9094661 9098547 9100711 9100712 LOTS NOT INCLUDED IN THE SCOPE OF THE RECALL ARE NOT AFFECTED. 3. DESCRIPTION OF ISSUE BD HAS CONFIRMED THAT THESE LOTS MAY EXHIBIT SEPARATION OF FRONT AND REAR BARRELS UPON ACTIVATION OF THE SAFETY FEATURE THAT RETRACTS THE NEEDLE. BARREL SEPARATION MAY RESULT IN LEAKAGE OF BLOOD OR AN EXPOSED NEEDLE, POTENTIALLY RESULTING IN EXPOSURE TO BLOOD BORNE PATHOGENS. BD PAS IDENTIFIED THIS ISSUE THRU THE BD COMPLAINT INVESTIGATION SYSTEM. 4. SUMMARY TABLE OF THE # OF COMPLAINTS AND MDRS IN SCOPE PER 806 # 2243072-07/22/2019-009-R DATED AUGUST 2, 2019, THERE WERE A TOTAL OF 33 COMPLAINTS AND 31 MDRS WITHIN SCOPE. COMPLAINT MDR (B)(4). 5. HHE SUMMARY THE HHE CONSIDERED THE POTENTIAL HAZARDS, HARMS, SEVERITIES, AND LIKELIHOOD OF OCCURRENCE RELATED TO THE PRODUCT, SPECIFICALLY THE REPORTED ISSUE REGARDING THE SAFETY MECHANISM NOT FUNCTIONING AS INTENDED. EXPOSURE TO BLOOD, INCLUDING BLOODBORNE PATHOGENS REPRESENTS A POTENTIAL FOR IMMEDIATE AND LONG-TERM HEALTH CONSEQUENCE TO THE HEALTHCARE WORKER. THIS WOULD BE IN THE PRESENCE OF A NEEDLESTICK INJURY TO THE HCW AFTER THE DEVICE WAS USED ON A PATIENT AS WELL AS BLOOD LEAKAGE FROM THE DEVICE WHICH COULD COME IN DIRECT CONTACT WITH THE UNPROTECTED SKIN (NOT FOLLOWING PROPER PPE PROCEDURE) OF THE HCW COLLECTING THE BLOOD. THE LIKELIHOOD OF HEALTH RISK FROM THE USAGE OF UNNECESSARY POST-EXPOSURE PROPHYLAXIS (PEP) AND LABORATORY TESTING DUE TO CONTAMINATED NEEDLE STICK INJURY OR EXPOSURE OF NON-INTACT SKIN IS CONSIDERED MINIMAL. THIS IS SUPPORTED BY DOCUMENTED LITERATURE THAT INDICATES GOOD EFFICACY OF PEP BEING OBSERVED WITH NO SEROCONVERSION. GENERAL AWARENESS FOR THE USE OF SHARPS AS WELL AS THE USE OF GENERAL SAFETY PRECAUTIONS IS KNOWN TO HELP DECREASE THE LIKELIHOOD OF EXPOSURE TO BLOODBORNE PATHOGENS. 6. INVESTIGATION SUMMARY ROOT CAUSE INVESTIGATION INDICATES A MIS-ALIGNMENT OF TOOLING TO LOAD AND ASSEMBLE COMPONENTS IN THE ASSEMBLY LINE. BD PAS HAS INITIATED CAPA 982194 TO FURTHER INVESTIGATE THIS ISSUE, AND CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED. 7. RECALL REFERENCE #, EITHER BD INTERNAL OR RES/Z# PLEASE REFERENCE BD RECALL # PAS-19-1567. H3 OTHER TEXT : SEE H.10

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER EXPERIENCED A DELAYED RETRACTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 368656 BATCH NO. 9081829 I WANTED TO MAKE YOU AWARE OF AN URGENT ISSUE WITH VACUTAINER ITEM# 368656 LOT# 9081829. THE SAFETY MECHANISM REPORTED TO FAIL BY THE NURSE USING THE ITEM. THE DEVICE ITSELF WAS NOT SAVED DUE TO THE CONTAMINATION RISK BUT THE PACKAGING WAS SAVED. THE CLINICAL TEAM ATTEMPTED TO DEACTIVATE THE SAFETY MECHANISM ON SEVERAL DEVICES IN THE SAME LOT AND WERE ABLE TO. WE HAVE PULLED AND SEQUESTERED THIS LOT. WE ARE ASKING THAT YOU PICK UP THE PRODUCT AND SEND SOME FOR QUALITY TESTING TO DETERMINE IF THERE IS A PRODUCT DEFECT IN THIS LOT OR WHAT MAY BE CAUSING THE ISSUE. WE HAVE A SIGNIFICANT AMOUNT OF PRODUCT THAT IS OF THE SAME LOT NUMBER THAT WE HAVE TO PULL FROM USE AND REPLACE.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE TYPE: FMI/JKA. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER EXPERIENCED A DELAYED RETRACTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 368656, BATCH NO. 9081829. I WANTED TO MAKE YOU AWARE OF AN URGENT ISSUE WITH VACUTAINER ITEM# 368656 LOT# 9081829. THE SAFETY MECHANISM REPORTED TO FAIL BY THE NURSE USING THE ITEM. THE DEVICE ITSELF WAS NOT SAVED DUE TO THE CONTAMINATION RISK BUT THE PACKAGING WAS SAVED. THE CLINICAL TEAM ATTEMPTED TO DEACTIVATE THE SAFETY MECHANISM ON SEVERAL DEVICES IN THE SAME LOT AND WERE ABLE TO. WE HAVE PULLED AND SEQUESTERED THIS LOT. WE ARE ASKING THAT YOU PICK UP THE PRODUCT AND SEND SOME FOR QUALITY TESTING TO DETERMINE IF THERE IS A PRODUCT DEFECT IN THIS LOT OR WHAT MAY BE CAUSING THE ISSUE. WE HAVE A SIGNIFICANT AMOUNT OF PRODUCT THAT IS OF THE SAME LOT NUMBER THAT WE HAVE TO PULL FROM USE AND REPLACE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459354 BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER BLOOD SPECIMEN COLLECTION DEVICE FMI BECTON, DICKINSON & CO., (BD) 9081829 50382903686563

Patients

Seq Age Sex Outcome Treatment
1 Other