FDA Adverse Event Injury Summary report: N

GUNTHER TULIP JUGULAR VENA CAVA FILTER SET

MDR report key: 8666197 · Received June 4, 2019

Report

Report Number
1820334-2019-01289
Event Type
Injury
Date Received
June 4, 2019
Report Date
August 21, 2019
Manufacturer
COOK INC
Product Code
DTK
UDI-DI
10827002330174
PMA / PMN Number
K072240
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: INVESTIGATION. INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: VENA CAVA PERFORATION, TILT, AND ANXIETY. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. VENA CAVA WALL PENETRATION/PERFORATION HAS BEEN REPORTED AND MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: TRAUMA TO ADJACENT STRUCTURES, VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. FILTER TILT HAS BEEN REPORTED. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER); AND (OR) A FAILED RETRIEVAL ATTEMPT. EXCESSIVE FILTER TILT MAY CONTRIBUTE TO DIFFICULT OR FAILED RETRIEVAL; VENA CAVA WALL PENETRATION/PERFORATION; AND (OR) RESULT IN LOSS OF FILTER EFFICIENCY. POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: UNACCEPTABLE FILTER TILT. UNKNOWN IF THE REPORTED ANXIETY/FEAR IS DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. NO RELEVANT NOTES ON NEITHER DEVICE OR LOT NUMBER. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE, OR UNCHANGED. THE FOLLOWING FIELDS WERE UPDATED PER ADDITIONAL INFORMATION RECEIVED: B5, B6, AND H6. INVESTIGATION THE INVESTIGATION WAS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. PER QUALITY ENGINEERING REVIEW, THE ADDITIONAL INFORMATION PROVIDED FOR THIS COMPLAINT DOES NOT CHANGE THE PREVIOUS INVESTIGATION CONCLUSION. THEREFORE, NO NEW INVESTIGATION ACTIVITIES WILL BE CONDUCTED AT THIS TIME. COOK WILL REOPEN THE INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. AS A RESULT, APPROPRIATE TERM/CODE NOT AVAILABLE (4316) WAS SELECTED FOR THE H6 CONCLUSION CODE. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AT THIS TIME.

Description of Event or Problem · 0

PER A REPORT FROM CT (COMPUTED TOMOGRAPHY) DATED (B)(6) 2017, "POSITIVE FOR CAVAL PERFORATION. SUPERIOR EXTENT OF IVC FILTER L1-L2 DISC SPACE. INFERIOR EXTENT L3-L4 VERTEBRAL BODY. A TOTAL OF FOUR PRONGS HAVE PERFORATED THE IVC SERIES 2 IMAGE 65. MAXIMUM DISTANCE PRONGS PERFORATED 5 MM SERIES 2 IMAGE 65. CORONAL IMAGES 7 DEGREE TILT RIGHT TO LEFT SERIES 602 IMAGE 48. SAGITTAL IMAGES 4 DEGREE TILT ANTERIOR POSTERIOR SERIES 601 IMAGE 80. DIAMETER OF IVC DIRECTLY ABOVE FILTER 31 MM X 23 MM SERIES 2 IMAGE 47."

Additional Manufacturer Narrative · 1

THE PREVIOUS MDR WAS SUBMITTED BY (FORMER MANUFACTURER) UNDER MANUFACTURER REPORT REFERENCE # 3002808486-2019-00596. OCCUPATION: NON-HEALTHCARE PROFESSIONAL. PMA/ 510K: K072240. (B)(4).

Description of Event or Problem · 1

ADDITIONAL INFORMATION REPORTS THAT THE PATIENT ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2012 VIA THE RIGHT INTERNAL JUGULAR VEIN BEFORE HIP/KNEE SURGERY AS A PRECAUTION. PATIENT IS ALLEGING TILT AND VENA CAVA PERFORATION. THE PATIENT FURTHER ALLEGES "I LIVE WITH THE ANXIETY OF HAVING THIS FILTER THAT COULD FAIL AT ANY TIME, BUT THAT CANNOT BE RETRIEVED WITHOUT A SERIOUS SURGERY. I LIVE WITH THE FEAR THAT MY FILTER HAS TILTED WITHIN MY VENA CAVA AND PERFORATED OUTSIDE OF IT''.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459723 GUNTHER TULIP JUGULAR VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK COOK INC G33017 3214571 10827002330174

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening