BATTERY,PAD-PAK-01
Report
- Report Number
- 3004123209-2019-00207
- Event Type
- Malfunction
- Date Received
- June 4, 2019
- Date of Event
- May 1, 2019
- Report Date
- July 30, 2019
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- P160008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER).
THE PAD-PAK WAS MANUFACTURED AS PART OF A LOT OF (B)(4) ON THE (B)(6)2017 . ALL PAD-PAK¿S WERE VOLTAGE TESTED ON THE (B)(6)2017 WITH ONE RECORDED FAIL. THE PAD-PAK WHICH FAILED WAS THEN STORED IN QUARANTINE. (B)(4) PAD-PAK UNITS WERE DISPATCHED TO ¿HST LLC¿ ON THE (B)(6)2017 . DAMAGE WAS OBSERVED ON THE LOCATOR PIN ON THE BATTERY TERMINAL SIDE OF THE RETURNED PAD-PAK (J0268 ¿(B)(6)2021).THIS DAMAGE IMPEDED THE INSERTION OF THE PAD-PAK WITHIN THE DEVICE AND RESULTED IN A DEVICE NOT POWERING ON.THE INVESTIGATION WAS UNABLE TO DETERMINE WHEN THE DAMAGE TO THE LOCATOR PIN OCCURRED BUT IF IT WAS PRESENT DURING THE INSTALLATION IT WOULD HAVE RESULTED IN THE REPORTED FAILURE. THE DAMAGE MAY HAVE BEEN CAUSED DURING AN INCORRECT INSTALLATION OF THE PAD-PAK. THE PAD-PAK WILL BE SCRAPPED AND REPLACED.
NEW REPLACEMENT PAD-PAK OUT OF BOX FAILED TO START DEVICE BUT OLD PAD PAK STILL POWERED UP DEVICE. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
NEW REPLACEMENT PAD-PAK OUT OF BOX FAILED TO START DEVICE BUT OLD PAD PAK STILL POWERED UP DEVICE.THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459412 | BATTERY,PAD-PAK-01 | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | J0268 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |