ANIMAS VIBE
Report
- Report Number
- 2531779-2019-03457
- Event Type
- Malfunction
- Date Received
- June 4, 2019
- Report Date
- June 2, 2019
- Manufacturer
- ANIMAS LLC
- Product Code
- OYC
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- 003
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05-AUG-2019 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP POWERED ON WITH A BLANK DISPLAY. THE COMPLAINT OF LINES THROUGH THE DISPLAY WAS NOT ABLE TO BE INVESTIGATED DUE TO A CRACKED DISPLAY WHICH CAUSED THE BLANK DISPLAY ISSUE. THE DAMAGED DISPLAY ISSUE WAS REPORTED ON ANIMAS MEDWATCH REPORT NUMBER 2531779-2019-03458. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (LINE THROUGH DISPLAY) ISSUE; A RED LINE THROUGH THE DISPLAY SCREEN. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 462041 | ANIMAS VIBE | INSULIN INFUSION PUMP | OYC | ANIMAS LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |