FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 8665960 · Received June 4, 2019

Report

Report Number
2531779-2019-03457
Event Type
Malfunction
Date Received
June 4, 2019
Report Date
June 2, 2019
Manufacturer
ANIMAS LLC
Product Code
OYC
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05-AUG-2019 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP POWERED ON WITH A BLANK DISPLAY. THE COMPLAINT OF LINES THROUGH THE DISPLAY WAS NOT ABLE TO BE INVESTIGATED DUE TO A CRACKED DISPLAY WHICH CAUSED THE BLANK DISPLAY ISSUE. THE DAMAGED DISPLAY ISSUE WAS REPORTED ON ANIMAS MEDWATCH REPORT NUMBER 2531779-2019-03458. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (LINE THROUGH DISPLAY) ISSUE; A RED LINE THROUGH THE DISPLAY SCREEN. THERE IS NO INDICATION THE ALLEGED PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462041 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS LLC

Patients

Seq Age Sex Outcome Treatment
1 43 YR