FDA Adverse Event Malfunction Summary report: N

FREESTYLE PRECISION NEO

MDR report key: 8665750 · Received June 4, 2019

Report

Report Number
2954323-2019-04395
Event Type
Malfunction
Date Received
June 4, 2019
Date of Event
May 30, 2019
Report Date
July 4, 2019
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
NBW
PMA / PMN Number
K140371
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. NOTE: THE DEVICE MANUFACTURER DATE FOR THE REPORTED METER SERIAL NUMBER IS UNKNOWN. THE DATE ENTERED IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT.

Additional Manufacturer Narrative · 1

THE RETURNED METER WAS INVESTIGATED WITH RETAINED TEST STRIPS. METER POWERED ON WITH BUTTON DEPRESSION AND TEST STRIP INSERTION. NO NEW ISSUES WERE OBSERVED. BLANK SCREEN WAS NOT OBSERVED. THE COMPLAINT WAS NOT CONFIRMED.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460822 FREESTYLE PRECISION NEO BLOOD GLUCOSE MONITORING SYSTEM NBW ABBOTT DIABETES CARE INC 71371-75

Patients

Seq Age Sex Outcome Treatment
1