PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL
Report
- Report Number
- 2032227-2019-11018
- Event Type
- Injury
- Date Received
- June 4, 2019
- Date of Event
- May 29, 2018
- Report Date
- July 16, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169507234
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
DEVICE PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED. MOTOR PASSED MOTOR TEST. (B)(4).
THE CUSTOMER'S FAMILY REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HYPERGLYCEMIA AND INSULIN PUMP RECEIVED MOTOR ERROR. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL AT TIME OF INCIDENT WAS UP TO 400 MG/D. THE CUSTOMER WAS TREATED WITH INSULIN PUMP, MANUAL INJECTION. THE CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER WAS ABLE TO COMPLETE INSULIN PUMP REWIND. CUSTOMER REPORTS THE DRIVE SUPPORT CAP APPEARS NORMAL. THE CUSTOMER WAS ADVISED TO DISCONTINUE THE USE OF INSULIN PUMP AND REVERT TO BACK UP. THE CUSTOMER DECLINED TROUBLESHOOTING. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS AND THE RESERVOIR WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 460808 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAL | B5723NALJ | 00643169507234 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR | Other | FRN-MMT-332-RSVR, UNOMED SET |