MENTOR SMOOTH ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2019-13130
- Event Type
- Injury
- Date Received
- June 3, 2019
- Date of Event
- August 2, 2018
- Report Date
- May 7, 2019
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001249
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED ON 7/30/2019 FOR THE FINISHED DEVICE NUMBER 5842160, AND NO NON-CONFORMANCES RELATED TO THE REPORTED COMPLAINT WERE IDENTIFIED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION ON 7/2/2019. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ADDITIONAL INFORMATION WAS RECEIVED WITH THE PRODUCT INVESTIGATION. THE LOT NUMBER IS 5842160. MENTOR IS AWARE THAT THE MANUFACTURED DATE OF THIS DEVICE IS AFTER THE REPORTED IMPLANT DATE, HOWEVER, MENTOR IS REPORTING THE INFORMATION MADE AVAILABLE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE IN THE FUTURE, THEN A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
THE INVESTIGATION OF THE RETURNED DEVICE WAS COMPLETED BY THE FAILURE ANALYSIS LAB ON 7/26/2019. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION PROVIDED, IT WAS REPORTED THAT THE PATIENT EXPERIENCED A DEFLATION OF THE IMPLANT. DURING EVALUATION OF THE SAMPLE IT WAS OBSERVED TO BE INTACT. LEAK TESTING WAS PERFORMED, AND IT REVEALED THREE TEARS ON THE POSTERIOR ASPECT, ALL MEASURING LESS THAN 0.1 CM, AND A LEAKAGE FROM THE VALVE. MICROSCOPIC EXAMINATION OF THE EDGES OF THE TEAR GAVE NO INDICATIONS AS TO CAUSE OF THE FAILURE. THE ANALYSIS WAS UNABLE TO DETERMINE THE ROOT CAUSE FOR THE LEAKING VALVE. EXCESSIVE MANIPULATION MAY HAVE CAUSED THE VALVE LEAK. THE INSTRUCTIONS FOR USE STATES ¿NOT TO MANIPULATE (I.E., SQUEEZE) THE VALVE EXCESSIVELY, WHICH MAY CAUSE VALVE LEAKAGE¿. HOWEVER, THIS CANNOT BE CONCLUSIVELY DETERMINED. REGARDING THE TEARS FOUND, POSSIBLE CAUSES, INCLUDE BUT ARE NOT LIMITED TO, THE FOLLOWING EVENTS: EXCESSIVE STRESSES OR MANIPULATIONS AS MAY OCCUR DURING NORMAL DAILY ROUTINES INCLUDING CUSTOMARY AND PURPOSEFUL TRAUMA TO THE BREAST. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. BASED ON THE INFORMATION REPORTED AND/OR THE PRODUCT INVESTIGATION, THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH THE MANUFACTURING PROCESS. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) YEAR-OLD PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH A MENTOR SMOOTH ROUND MODERATE PROFILE 300CC SALINE PROSTHESIS EXPERIENCED RIGHT SIDED DEFLATION POST PROCEDURE. THE DEFLATION DIAGNOSED VISUALLY AND PALPATING THE BREAST. DEVICE REPLACEMENT WITH ANOTHER MENTOR SMOOTH ROUND MODERATE PROFILE 300CC SALINE PROSTHESIS WAS PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 458159 | MENTOR SMOOTH ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 5842160 | 00081317001249 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Required Intervention |