STEALTHSTATION® S7
Report
- Report Number
- 1723170-2019-02887
- Event Type
- Malfunction
- Date Received
- June 3, 2019
- Report Date
- August 8, 2019
- Manufacturer
- MEDTRONIC NAVIGATION, INC
- Product Code
- HAW
- UDI-DI
- 00643169505742
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CI
- Reporter Occupation
- OTHER
Narratives
SUMMARY: ORIGINAL PROBLEM WAS SCU FUSES FAIL 2 TIMES. SECOND PROBLEM WAS IMPORTATION OF CRANIAL PATIENT FROM OARM.(3.0.1 KNOWN PROBLEMS); RESOLVED: YES SOFTWARE: FAIL; FAILURES: IMPORT EXAM; SUMMARY: ORIGINAL PROBLEM WAS SCU FUSES FAIL 2 TIMES. SECOND PROBLEM WAS IMPORTATION OF CRANIAL PATIENT FROM OARM.(3.0.1 KNOWN PROBLEMS); MNAV ACTION/RESOLUTION: DEFINITIVE REPLACEMENT OF THE FOLLOW SPARE PARTS: - 9733582 LOT150127 BY LOT161221 - 9733575, LOT150720 BY LOT120820 - 9733437, S/N: (B)(4) - 9735586 CRANIAL 3.0.2. SUMMARY: CAMERA LED OFF. SCU LED OFF. FUSES UN SCU OPEN. THIS FUSES WAS REPLACED AND ALL SYSTEM WORKED AGAIN. MNAV ACTION/RESOLUTION: 2 FUSES 2A 250V INSIDE THE SCU (9733438) WERE REPLACED. THIS ALLOWED TO USE THE EQUIPMENT TEMPORARILY. IS RECOMMENDED THE REPLACEMENT OF SCU, PSU AND CABLE. OTHER RELEVANT DEVICE(S) ARE: MODEL: 9733581. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE POSITIONING SENSOR UNIT (PSU) WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE PSU WAS FOUND TO BE FULLY FUNCTIONAL WITH NO PROBLEM FOUND. THE REPORTED EVENT COULD NOT BE DUPLICATED BY MEDTRONIC PERSONNEL. THE PSU TO POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) CABLE WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE CABLE WAS FOUND TO BE FULLY FUNCTIONAL WITH NO PROBLEM FOUND. THE REPORTED EVENT COULD NOT BE DUPLICATED BY MEDTRONIC PERSONNEL. THE EXTERNAL PSU TO SCU CABLE WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE CABLE WAS FOUND TO BE FULLY FUNCTIONAL WITH NO PROBLEM FOUND. THE REPORTED EVENT COULD NOT BE DUPLICATED BY MEDTRONIC PERSONNEL. THE SCU WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. TESTING FOUND THAT THE SCU WOULD NOT POWER ON. THE LINE FILTER FUSES WERE NOTED TO HAVE ONE OPEN FUSE. THE HARDWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS RELATED TO A HARDWARE ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE POLARIS SPECTRA SYSTEM CONTROL UNIT (SCU) WILL NOT POWER ON. THERE WAS NO PATIENT PRESENT AT THE TIME OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457929 | STEALTHSTATION® S7 | INSTRUMENT, STEREOTAXIC | HAW | MEDTRONIC NAVIGATION, INC | 9733858 | 00643169505742 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |