FDA Adverse Event Malfunction Summary report: N

K-WIRE SS 1.6X127MM .062 ST

MDR report key: 8663956 · Received June 3, 2019

Report

Report Number
0001825034-2019-02379
Event Type
Malfunction
Date Received
June 3, 2019
Date of Event
May 14, 2019
Report Date
September 27, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
HXC
PMA / PMN Number
EXEMPT
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL EVALUATION OF THE RETURNED PRODUCT IDENTIFIED THAT THERE IS HAIR-LIKE DEBRIS INSIDE THE STERILE PACKAGE. FTIR ANALYSIS CONDUCTED ON THE FOREIGN MATERIAL IN THE PACKAGE IDENTIFIED THAT IT IS CONSISTENT WITH THE FTIR SPECTRA OF BIOLOGICAL MATERIAL (LIKELY A HAIR). DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. THE PRODUCT WAS LIKELY NON-CONFORMING WHEN IT LEFT ZIMMER BIOMET CONTROL. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO THE OPERATOR NOT FOLLOWING INSTRUCTIONS DURING MANUFACTURING. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). REPORT SOURCE: (B)(6). CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INSPECTION, HAIR LIKE DEBRIS WAS FOUND IN THE STERILE PACKAGE. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.