ILS 25MM, CURVED
Report
- Report Number
- 3005075853-2019-19460
- Event Type
- Malfunction
- Date Received
- June 3, 2019
- Date of Event
- May 22, 2019
- Report Date
- May 22, 2019
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- UDI-DI
- 10705036003458
- PMA / PMN Number
- K983536
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). BATCH # N54U27. INVESTIGATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE CDH25A DEVICE ARRIVED WITH NO APPARENT DAMAGE. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE, AND THE STAPLES MEET THE STAPLE FORM RELEASE CRITERIA. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DIFFICULTIES NOTED. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED.
(B)(4). BATCH # UNK. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER AND NO NON-CONFORMANCES WERE IDENTIFIED.
IT WAS REPORTED THAT DURING A COLECTOMY PROCEDURE, THE STAPLER DID NOT PROCESS THE STAPLING AT THE TIME OF THE SHOT. THERE WAS NO REPORTED ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 458466 | ILS 25MM, CURVED | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | N4LJ67 | 10705036003458 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |