FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE

MDR report key: 8663532 · Received June 3, 2019

Report

Report Number
1917413-2019-01560
Event Type
Malfunction
Date Received
June 3, 2019
Date of Event
May 17, 2019
Report Date
May 20, 2019
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903627603
PMA / PMN Number
K911731
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. INVESTIGATION CONCLUSION: AS NO SAMPLES OR PHOTOS WERE RECEIVED FOR EVALUATION, THE CUSTOMER'S INDICATED FAILURE MODE WAS NOT OBSERVED BY BD. ROOT CAUSE DESCRIPTION: AS THERE WAS NO SAMPLE OR PHOTO AVAILABLE FOR EVALUATION, A ROOT CAUSE COULD NOT BE DETERMINED. RATIONALE: COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT TUBES WERE UNDER FILLING WITH A BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE. THIS OCCURRED ON 120 SEPARATE OCCASIONS BUT THE DATE/TIME AND OR PATIENT INFORMATION IS UNKNOWN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 362760, BATCH NO. 8152662. IT WAS REPORTED THE TUBES WERE UNDERFILLING. "THESE TUBES AREN'T COLLECTING ANY BLOOD DURING BLOOD DRAW. TUBE WORKS WHEN ANOTHER LOT IS USED SO IT'S REALLY JUST THIS LOT. WE HAVE NOTICED 2 WHOLE CASES OF TUBES (120) NOT FILLING. IT'S BEEN HAPPENING RECENTLY BUT I DON'T HAVE A SPECIFIC DATE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458422 BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 8152662 50382903627603

Patients

Seq Age Sex Outcome Treatment
1 Other