FDA Adverse Event Malfunction Summary report: N

BARDEX® I.C. TEMPERATURE SENSING FOLEY CATHETER

MDR report key: 8663135 · Received June 3, 2019

Report

Report Number
1018233-2019-02867
Event Type
Malfunction
Date Received
June 3, 2019
Date of Event
May 10, 2019
Report Date
July 12, 2019
Manufacturer
C.R. BARD, INC. (COVINGTON) -1018233
Product Code
MJC
PMA / PMN Number
K910318
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS UNCONFIRMED. THE EVALUATION FOUND THE RETURNED SAMPLE MET SPECIFICATION. OBSERVED READING AS USUAL ON THE RETURNED SAMPLE. DISSECTED THE CATHETER AND FOUND THE WIRE WAS IN GOOD CONDITION. THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "[CONTRAINDICATIONS] 1. METHOD FOR USE: (1) DO NOT REUSE. (2) DO NOT RESTERILIZE. (3) THIS DEVICE CONTAINS 10% POVIDONE-IODINE. FOR PATIENTS WITH PAST HISTORY OF ALLERGIC HYPERSENSITIVITY TO POVIDONE-IODINE OR IODINE, CONSIDER USING ALTERNATIVE DISINFECTANTS. (4) BE CAREFUL THAT THE CATHETER IS NOT EXPOSED TO OINTMENTS, CONTRAST MEDIUM OR OIL-BASED LUBRICANTS (INCLUDING VEGETABLE OILS SUCH AS OLIVE OIL, MINERAL OILS SUCH AS WHITE PETROLATUM AND ANIMAL OILS). [THEY MAY DAMAGE THE DEVICE AND MAY BURST BALLOON.] (5) DO NOT HOLD THE DEVICE WITH FORCEPS, ETC. AVOID CONTACT WITH ANY BLADES OR SHARP-EDGED INSTRUMENTS. [CATHETER DAMAGE MAY CAUSE BALLOON RUPTURE AND ACCIDENTAL BALLOON REMOVAL OR FAILURE TO DEFLATE OR REMOVE THE BALLOON.] 2. APPLICABLE PATIENTS (1)PATIENTS WHO ARE OR HAVE BEEN ALLERGIC TO NATURAL RUBBERLATEX (2)PATIENTS WITH KNOWN ALLERGY TO SILVER COATED CATHETER". SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CATHETER DID NOT INDICATE TEMPERATURE PROPERLY. THE DISPLAYED TEMPERATURE WAS AROUND 3 DEGREES LOWER THAN THE CORRECT TEMPERATURE OF THE PATIENT.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CATHETER DID NOT INDICATE TEMPERATURE PROPERLY. THE DISPLAYED TEMPERATURE WAS AROUND 3 DEGREES LOWER THAN THE CORRECT TEMPERATURE OF THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
454943 BARDEX® I.C. TEMPERATURE SENSING FOLEY CATHETER FOLEY CATHETER (LATEX) MJC C.R. BARD, INC. (COVINGTON) -1018233 8PH040

Patients

Seq Age Sex Outcome Treatment
1