EQUINOXE
Report
- Report Number
- 1038671-2019-00322
- Event Type
- Injury
- Date Received
- June 3, 2019
- Date of Event
- May 22, 2019
- Report Date
- April 7, 2020
- Manufacturer
- EXACTECH, INC.
- Product Code
- KWT
- UDI-DI
- 10885862086389
- PMA / PMN Number
- K063569
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
(B2) OUTCOMES ATTRIBUTED TO ADVERSE EVENT: ADDED CHECK FOR HOSPITALIZATION - INITIAL OR PROLONGED. (D6) IF IMPLANTED, GIVE DATE: (B)(6) 2018. (E3) OCCUPATION: PHYSICIAN. (G5) PMA/510(K)NUMBER: K063569. (H4) DEVICE MANUFACTURE DATE: 03-OCT-2018. (H3) THE REVISION REPORTED WAS LIKELY THE RESULT OF NOT PROPERLY ALIGNING THE GLENOSPHERE RELATIVE TO THE GLENOID BASEPLATE DURING THE INITIAL SURGERY. THIS MISALIGNMENT LIKELY PROHIBITED THE GLENOSPHERE FROM FULLY SEATING AND THE GLENOSPHERE LOCKING SCREW FROM FULLY ENGAGING THE GLENOID BASEPLATE, RESULTING IN DISASSOCIATION OF THE GLENOSPHERE. HOWEVER, THIS CANNOT BE CONFIRMED AS THE EXPLANTS WERE NOT AVAILABLE FOR EVALUATION. (H6) EVALUATION CODES: 3165, 2907. SECTION H11: THE FOLLOWING SECTIONS HAVE CORRECTED INFORMATION: (SECTION F) PLEASE DISREGARD F6 AND F8. THESE WERE ENTERED IN ERROR. (G4) INITIAL AWARENESS DATE IN INITIAL SUBMISSION SHOULD HAVE BEEN 22-MAY-2019 (H5) LABELED FOR SINGLE USE?: YES.
PENDING EVALUATION. CONCOMITANT DEVICES (2): EQ REV LOCKING SCREW (CN: 320-15-05; SN: (B)(4)). EQUINOXE REVERSE 38MM HUMERAL LINER +0 (CN: 320-38-00; SN: (B)(4)).
PRIMARY SURGERY: (B)(6) 2018. THE GLENOSPHERE WAS FOUND UNCOUPLED ON THE X-RAY AT A 6 MONTH FOLLOW UP WITH THE PATIENT. PATIENT WAS ASYMPTOMATIC, NO PAIN AND WITH FULL FUNCTIONING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456800 | EQUINOXE | EQUINOXE REVERSE 38MM GLENOSPHERE | KWT | EXACTECH, INC. | 320-01-38 | 10885862086389 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |