PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Report
- Report Number
- 2032227-2019-10715
- Event Type
- Malfunction
- Date Received
- June 3, 2019
- Date of Event
- February 27, 2019
- Report Date
- May 20, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 00763000190460
- Removal / Correction Number
- Z-0956-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0956-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP WAS RECEIVED WITH A BLANK DISPLAY DUE TO MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY. UNABLE TO PERFORM THE SELF TEST, DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, AND ACTIVE CURRENT MEASUREMENT TEST OR VERIFY CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM DUE TO BLANK DISPLAY. MOISTURE DAMAGE WAS ALSO FOUND ON THE MOTOR AND FORCE SENSOR DURING VISUAL INSPECTION. THE PUMP WAS RECEIVED WITH CRACKED BEHIND THE PUMP AT THE CORNER OF THE BELT CLIP RAILS, PARTIALLY BROKEN RESERVOIR TUBE LIP, MISSING RETAINER AND MISSING RESERVOIR TUBE O-RING. A TEST RESERVOIR WAS INSTALLED AND DID NOT LOCK IN PLACE DUE TO MISSING RETAINER.
THE CUSTOMER¿S MOTHER REPORTED THAT THE INSULIN PUMP RECEIVED AN OPEN BOOK IMAGE ON THE SCREEN OF INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE VALUE WAS UNKNOWN. CUSTOMER REPORTS THEY RECEIVED THE OPEN BOOK IMAGE ON THE INSULIN PUMP SCREEN. CUSTOMER REPORTS THERE WAS FLUID IN THE BATTERY COMPARTMENT. CUSTOMER STATES O-RING WAS IN PLACE. CUSTOMER STATES O-RING WAS FREE OF DEBRIS. CUSTOMER STATES BATTERY CAP WAS NOT DAMAGED. THE CUSTOMER ALSO REPORTED THAT THE RESERVOIR LIP RING WAS DAMAGED. THE CUSTOMER ALSO REPORTED THAT THE RESERVOIR WAS ABLE TO LOCK IN PLACE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE CUSTOMER WAS ASSISTED IN TROUBLESHOOTING. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457796 | PUMP MMT-1780KPK 670G PATHWAY BLACK MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KPK | HG2TNEZ | 00763000190460 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR | Male |