FDA Adverse Event
Malfunction
Summary report: N
ARCHITECT I2000SR ANALYZER
MDR report key: 866163
·
Received April 18, 2007
Report
- Report Number
- 1628664-2007-00106
- Event Type
- Malfunction
- Date Received
- April 18, 2007
- Date of Event
- February 21, 2007
- Report Date
- March 19, 2007
- Manufacturer
- ABBOTT LABORATORIES, INC
- Product Code
- JJE
- PMA / PMN Number
- k983212
- Removal / Correction Number
- 1628664-04112007-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED.
Description of Event or Problem · 1
MICROBIAL CONTAMINATION WITHIN THE ARCHITECT SYSTEM (ARCHITECT I2000/I2000 SR PROCESSING MODULES AND ARCHITECT WASH BUFFER) THROUGH GENERATION OF FOLATE-LIKE BY-PRODUCTS (MICROBIAL FOLATE) MAY CAUSE ARCHITECT FOLATE CALIBRATION FAILURES AND SHIFTS IN ARCHITECT FOLATE RESULTS (QC AND PT RESULTS).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT I2000SR ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT LABORATORIES, INC | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO |