2.7 DEGREE ANGLED SAGITAL SAW
Report
- Report Number
- 3005985723-2019-00401
- Event Type
- Malfunction
- Date Received
- May 31, 2019
- Date of Event
- May 7, 2019
- Report Date
- February 24, 2026
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- UDI-DI
- 00848486032128
- PMA / PMN Number
- K143752
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
REPORTED EVENT: IT WAS REPORTED THAT THE SAW ATTACHMENT COULD NOT HOLD SAW BLADE. PRODUCT EVALUATION AND RESULTS: THE PRODUCT WAS NOT EVALUATED AS THE PRODUCT WAS UNAVAILABLE FOR INSPECTION. PRODUCT HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED UNDER LOT NO 35010818 AND (B)(4) OUT OF (B)(4) ACCEPTED INTO FINAL STOCK ON 09/23/2018. REVIEW OF QT 18-09-0113 REVEALED THAT THE NON CONFORMANCE WAS NOT RELATED TO THE FAILURE ALLEGED IN THE COMPLAINTS, (B)(4) OUT OF (B)(4) WERE ACCEPTED INTO FINAL STOCK ON 08/28/2019. REVIEW OF QT 18-09-0113 REVEALED THAT THE NON CONFORMANCE WAS NOT RELATED TO THE FAILURE ALLEGED IN THE COMPLAINT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 212480, LOT NUMBER 35010818 SHOWS NO ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: THE FAILURE COULD NOT BE DETERMINED AS THE PRODUCT WAS NOT AVAILABLE FOR INSPECTION. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED. IF ADDITIONAL INFORMATION IS RECEIVED THEN THE COMPLAINT WILL BE REOPENED. CORRECTIVE ACTION/PREVENTIVE ACTION: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN NO NC AND CAPA ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. H3 OTHER TEXT : PRODUCT WAS NOT AVAILABLE FOR EVALUATION.
WHEN CUTTING BONE USING THE ANGLED SAW BLADE ATTACHMENT, THE BLADE KEPT SPITTING OUT, EVEN AFTER SCRUB TECH TIGHTENED DOWN THE BLADE USING THE HEX KEY. OPENED ANOTHER TRAY, PULLED OUT ANOTHER ANGLED SAW ATTACHMENT, AND THAT ONE WORKED APPROPRIATELY. AFTER CASE EXAMINED THE ATTACHMENT, AND NOTICED THAT THE RATCHETING NOISE SOUNDED SLIGHTLY DIFFERENT AND THAT YOU COULD OVER TORQUE THE TIGHTENING FUNCTION. SURGICAL DELAY APPROXIMATELY 1-2 MINUTES. CASE TYPE: TKA.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
WHEN CUTTING BONE USING THE ANGLED SAW BLADE ATTACHMENT, THE BLADE KEPT SPITTING OUT, EVEN AFTER SCRUB TECH TIGHTENED DOWN THE BLADE USING THE HEX KEY. OPENED ANOTHER TRAY, PULLED OUT ANOTHER ANGLED SAW ATTACHMENT, AND THAT ONE WORKED APPROPRIATELY. AFTER CASE EXAMINED THE ATTACHMENT, AND NOTICED THAT THE RATCHETING NOISE SOUNDED SLIGHTLY DIFFERENT AND THAT YOU COULD OVER TORQUE THE TIGHTENING FUNCTION. SURGICAL DELAY APPROXIMATELY 1-2 MINUTES. CASE TYPE: TKA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 453985 | 2.7 DEGREE ANGLED SAGITAL SAW | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 3503805 / 35010818 | 00848486032128 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |