FDA Adverse Event Malfunction Summary report: N

SPINNING SPIROS® CLOSED MALE LUER, PURPLE CAP

MDR report key: 8655979 · Received May 30, 2019

Report

Report Number
9617594-2019-00151
Event Type
Malfunction
Date Received
May 30, 2019
Date of Event
May 8, 2019
Report Date
May 9, 2019
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
FPA
PMA / PMN Number
K082806
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THREE USED LIST # CH2000S-PC, SPINNING SPIROS® CLOSED MALE LUER, PURPLE CAP; SUSPECTED LOT #S 3972031 AND 3963910 WERE RECEIVED FOR EVALUATION. EACH OF THE USED CH2000S-PC SPINNING SPIROS ATTACHED TO VARIOUS SYRINGES WERE FUNCTIONALLY TESTED AS RETURNED. NO LEAKAGE WAS OBSERVED AT ANY CONNECTION OR LOCATION WITHIN THE CH2000S-PC SPINNING SPIROS ASSEMBLY. EACH OF THE USED CH2000S-PC SPINNING SPIROS WERE PRESSURE LEAK TESTED AND EACH PRESSURE LEAK EXPECTATIONS OUTLINED IN THE PRODUCT PERFORMANCE SPECIFICATION. THE COMPLAINT WAS UNABLE TO BE CONFIRMED. THE SUSPECTED LOT #S 3972031 AND 3963910 WERE REVIEWED AND THERE WERE NO RELEVANT NON-CONFORMANCES FOUND.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED FOR TESTING AND INVESTIGATION; HOWEVER, THE INVESTIGATION IS NOT COMPLETE. PLOTS: SUSPECTED LOT NUMBERS: 3972031 (EXPIRY DATE 2024-02-01, MFR DATE 2019-02-01) AND 3963910 (EXPIRY DATE 2024-01-01; MFR DATE 2019-01-01).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT UPON CONNECTION OF TWO SPIROS CONNECTORS, THERE WAS VISIBLE MEDICATION OUTSIDE OF THE INTERNAL TUBE OF THE DEVICE. THE SPIROS WAS CONNECTED TO AN UNSPECIFIED VIAL ADAPTER. THERE WAS NO ADVERSE EVENT OR MEDICAL INTERVENTION REQUIRED, HOWEVER, UNPROTECTED CHEMO EXPOSURE WAS REPORTED. THIS REPORT CAPTURES 1 OF 2 SPIROS INVOLVED IN THE INCIDENT.