MEDTRONIC NAVIGATION
Report
- Report Number
- 1723170-2019-02795
- Event Type
- Injury
- Date Received
- May 30, 2019
- Date of Event
- May 1, 2015
- Report Date
- May 30, 2019
- Manufacturer
- MEDTRONIC NAVIGATION, INC
- Product Code
- HAW
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
Narratives
AGE OR DATE OF BIRTH: THE AVERAGE AGE OF THE 24 PATIENTS IS 62.2 YEARS. SEX: THERE WERE 3 MALE PATIENTS AND 21 FEMALE PATIENTS. WEIGHT: PATIENT WEIGHT NOT AVAILABLE FROM THE SITE. DATE OF EVENT: EVENT DATE IS APPROXIMATED AS IT WAS REPORTED THAT THE SURGERIES OCCURRED BETWEEN MAY 2015 AND OCTOBER 2017. DESCRIPTION OF PROBLEM OR EVENT: THERE WAS NO ALLEGATION OF MALFUNCTION OF THE SYSTEM. LOT #, SERIAL #: DEVICE LOT NUMBER, OR SERIAL NUMBER AND MODEL NUMBER WERE UNAVAILABLE. UDI #: UDI NOT AVAILABLE FOR THIS SYSTEM AT TIME OF FILING. PMA/510K: THE PMA/ 510 (K) NUMBER IS UNAVAILABLE SINCE THE MODEL NUMBER IS UNKNOWN. DEVICE EVALUATED BY MFR: THE DEVICE WAS NOT RETURNED, SO NO ANALYSIS WAS CONDUCTED. MANUFACTURE DATE: DEVICE MANUFACTURING DATE IS DEPENDENT ON LOT NUMBER/SERIAL NUMBER, THEREFORE, UNAVAILABLE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
CITATION: WORLD NEUROSURG. (2019) 122: E1588-E1591. BACKGROUND: SIGNIFICANT PROGRESS IN HARDWARE AND SURGICAL TECHNIQUES FOR SACROILIAC JOINT (SIJ) FUSION SURGERIES HAS FACILITATED SAFER AND MORE EFFICACIOUS PROCEDURES FOR PATIENTS. TRIANGULAR-SHAPED IMPLANTS FOR SIJ FUSIONS ARE THE MOST-STUDIED DEVICES AND HAVE DEMONSTRATED GOOD SHORT-TERM AND LONG-TERM CLINICAL OUTCOMES. REPORTS ON CYLINDRICAL THREADED IMPLANTS ARE VERY LIMITED. OWING TO BIOMECHANICAL DIFFERENCES IN THE IMPLANTS AND THE SURGICAL TECHNIQUES REQUIRED FOR THEIR PLACEMENT, PREVIOUSLY REPORTED RESULTS MAY NOT BE APPLICABLE TO CYLINDRICAL THREADED IMPLANTS. THE AIM OF THIS STUDY WAS TO REPORT PRELIMINARY CLINICAL EXPERIENCE WITH MINIMALLY INVASIVE SIJ FUSION USING INTRAOPERATIVE STEREOTACTIC NAVIGATION AND THE RIALTO SI FUSION SYSTEM. METHODS: WE RETROSPECTIVELY REVIEWED 24 PATIENTS WHO UNDERWENT SIJ FUSIONS BETWEEN MAY 2015 AND OCTOBER 2017 PERFORMED BY A SINGLE SURGEON. RESULTS: MEAN TOTAL SATISFACTION SCORE WAS 89.0% 27.6%. A STATISTICALLY SIGNIFICANT REDUCTION (P [ 0.0028) IN LOW BACK PAIN SCORES WAS NOTED FROM AN AVERAGE BASELINE SCORE OF 6.6 2.4 TO 3.7 3.3 POSTOPERATIVELY. LEG PAIN SCORES DECREASED FROM 4.8 3.8 TO 1.5 2.9 (P [ 0.0034). MEAN SURGICAL TIME WAS 53.0 13.9 MINUTES. IT TOOK SIGNIFICANTLY LONGER (P [ 0.0089) TO PERFORM THE INITIAL 13 CASES (59.9 15.2 MINUTES) COMPARED WITH SUBSEQUENT CASES (45.4 7.3 MINUTES). ESTIMATED BLOOD LOSS WAS MINIMAL (10.4 5.2 ML). CONCLUSIONS: MINIMALLY INVASIVE SI JOINT FUSION USING CYLINDRICAL THREADED IMPLANTS CAN BE SAFELY PERFORMED WITH MINIMAL MORBIDITY AND GOOD CLINICAL OUTCOMES. ADVERSE EVENTS: 2 PATIENTS HAD SYMPTOMATIC SUBCUTANEOUS HEMATOMAS, WHICH RESOLVED SPONTANEOUSLY. 2 PATIENTS HAD SUPERFICIAL WOUND INFECTIONS TREATED WITH ANTIBIOTICS. ONE PATIENT DEVELOPED AN OSTEOPHYTE ON THE LATERAL ASPECT OF THE IMPLANT. BECAUSE OF SYMPTOMATIC PAIN LOCALIZED TO THAT AREA, AN OSTEOPHYTECTOMY WAS PERFORMED 1 YEAR AFTER THE INDEX SURGERY WITH AN IMPROVEMENT IN PATIENT SYMPTOMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449721 | MEDTRONIC NAVIGATION | INSTRUMENT, STEREOTAXIC | HAW | MEDTRONIC NAVIGATION, INC | UNK_NAV_SYS | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other |