FDA Adverse Event
Other
Summary report: N
CORSON NEEDLE FOR ESU DISCRETION
MDR report key: 86551
·
Received August 2, 1994
Report
- Report Number
- 86551
- Event Type
- Other
- Date Received
- August 2, 1994
- Date of Event
- March 8, 1994
- Report Date
- March 11, 1994
- Manufacturer
- KARL STORZ ENDOSCOPY
- Product Code
- GEI
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- MD, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
ELECTROSURGICAL NEEDLE; INSULATION BROKE DOWN AT BALL, CAUSING THE HOLDER TO BECOME ENERGIZED SHORTLY AFTER THE START OF CASE. SURGEON RECEIVED BURN ON HAND THROUGH GLOVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORSON NEEDLE FOR ESU DISCRETION | ESU PROBE | GEI | KARL STORZ ENDOSCOPY | * | 6684 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other | VALLEY LAB ESU FORCE 4 |