FDA Adverse Event Malfunction Summary report: N

PEN II MYLAN 3.0ML

MDR report key: 8651719 · Received May 29, 2019

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
2243072-2019-01036
Event Type
Malfunction
Date Received
May 29, 2019
Date of Event
May 10, 2019
Report Date
June 24, 2019
Manufacturer
BECTON DICKINSON
Product Code
FMI
PMA / PMN Number
UNKNOWN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: TWO (2) SAMPLES WERE PROVIDED TO BD MEDICAL ¿ PHARMACEUTICAL SYSTEM (BDM-PS) FOR ANALYSIS. BDM-PS PERFORMED A BATCH HISTORY RECORD¿S REVIEW (BHR) INCLUDING A REVIEW OF ALL DATA COLLECTED DURING IN PROCESS AND QUALITY INSPECTIONS. THE BATCH INVOLVED IN THIS COMPLAINT MEETS ALL ACCEPTABLE QUALITY LEVELS (AQL¿S), WAS MANUFACTURED AND RELEASED ACCORDING TO APPLICABLE PROCEDURES AND SPECIFICATIONS. INITIAL EVALUATION REVEALED NO VISIBLE DAMAGE TO THE PENS. THE PENS WERE TESTED FOR VOLUMETRIC DOSE ACCURACY PER IT1172. ALL DOSES WERE WITHIN SPECIFICATION, AND THE PENS FUNCTIONED AS INTENDED. AN INJECTION SEQUENCE WAS EXECUTED USING A 29G X 12.7MM BD PEN NEEDLE AND CARTRIDGE. BASED ON INVESTIGATION CONCLUSION, BDM-PS WAS NOT ABLE TO CONFIRM THE SYMPTOM PERCEIVED BY CUSTOMER OR CORRELATE THIS SYMPTOM WITH A POTENTIAL CAUSE LINKED TO BD PROCESS.

Additional Manufacturer Narrative · 0

THE CHANGES ARE AS FOLLOWS: THE MANUFACTURING LOCATION FOR THIS PRODUCT IS HDQ. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN (B)(6) HAS BEEN LISTED IN AND THE (B)(6) FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. MEDICAL DEVICE BRAND NAME: PEN II MYLAN 3.0ML. MEDICAL DEVICE TYPE: FMI. COMMON DEVICE NAME: PEN. MEDICAL DEVICE MANUFACTURER: BECTON DICKINSON. MEDICAL DEVICE CATALOG #: 47314378. MEDICAL DEVICE LOT #: 16218001. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2016-08-18. UNIQUE IDENTIFIER (UDI) #: (B)(4). MANUFACTURING LOCATION: BECTON DICKINSON.

Description of Event or Problem · 0

IT WAS REPORTED THAT NO MEDICATION DELIVERY OCCURRED WITH A PEN II MYLAN 3.0ML. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE CONSUMER STATED THAT HE DIALED THE SETTING TO 0.3 AND PUSHED THE INJECTION BUTTON DOWN TO INJECT, BUT NO MEDICATION CAME OUT." 2 OCCURRENCES REPORTED, DATE/TIME AND PATIENT INFORMATION WERE NOT GIVEN.

Description of Event or Problem · 0

IT WAS REPORTED THAT NO MEDICATION DELIVERY OCCURRED WITH A PEN II MYLAN 3.0ML. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "THE CONSUMER STATED THAT HE DIALED THE SETTING TO 0.3 AND PUSHED THE INJECTION BUTTON DOWN TO INJECT, BUT NO MEDICATION CAME OUT." 2 OCCURRENCES REPORTED, DATE/TIME AND PATIENT INFORMATION WERE NOT GIVEN.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NO MEDICATION DELIVERY OCCURRED WITH A UNSPECIFIED BD¿ PEN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "THE CONSUMER STATED THAT HE DIALED THE SETTING TO 0.3 AND PUSHED THE INJECTION BUTTON DOWN TO INJECT, BUT NO MEDICATION CAME OUT." 2 OCCURRENCES REPORTED, DATE/TIME AND PATIENT INFORMATION WERE NOT GIVEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
444189 PEN II MYLAN 3.0ML PEN FMI BECTON DICKINSON 16218001

Patients

Seq Age Sex Outcome Treatment
1 Other