INRATIO
Report
- Report Number
- 2954730-2007-00265
- Event Type
- Malfunction
- Date Received
- June 8, 2007
- Date of Event
- May 15, 2007
- Report Date
- June 8, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: 2007, INTRATIO 4.6 (PT#1), LAB 3.3, MEAN 3.95 CONFIDENCE LIMITS 2.3-5.7. DATE: THE SAME DAY, INTRATIO 4.7 (PT#2), LAB 3.6, MEAN 4.15, CONFIDENCE LIMITS 2.4-6.1. DATE: THE NEXT DAY, INTRATIO 2.4 (PT#1), LAB 2.3, MEAN 2.35, CONFIDENCE LIMITS 1.6-3.4. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE FURTHER TESTING IS NOT REQUIRED AT THIS TIME. TS UPDATED THIS CASE IN 2007. DATE: THE SAME DAY, INTRATIO 2.0 (PT#1), LAB 1.9, MEAN 1.95, CONFIDENCE LIMITS 1.3-2.7. DATE: THE SAME DAY, INTRATIO 3.0 (PT#2), LAB 2.5, MEAN 2.75, CONFIDENCE LIMITS 1.7-3.8. DATE: THE SAME DAY, INTRATIO 3.0 (PT#3), LAB 2.7, MEAN 2.85, CONFIDENCE LIMITS 1.8-4.2. DATE: THE SAME DAY, INTRATIO 2.4 (PT#4), LAB 2.3, MEAN 2.35, CONFIDENCE LIMITS 1.6-3.4. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE FURT.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE 05/14/07, INTRATIO 4.6 (PT#1), LAB 3.3, DATE 2007, INTRATIO 4.7 (PT#2), LAB 3.6, DATE: THE NEXT DAY, INTRATIO 2.4 (PT#1), LAB 2.3, DATE: ONE DAY LATER, INTRATIO 2.0 (PT#1), LAB 1.9, DATE: THE SAME DAY, INTRATIO 3.0 (PT#2), LAB 2.5, DATE: THE SAME DAY, INTRATIO 3.0 (PT#3), LAB 2.7, DATE: THE SAME DAY, INTRATIO 2.4 (PT#4), LAB 2.3.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 070070 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |