FDA Adverse Event Summary report: N

QUICK CHECK ONE

MDR report key: 86487 · Received April 24, 1997

Report

Report Number
2027242-1997-90133
Date Received
April 24, 1997
Date of Event
March 26, 1997
Report Date
March 26, 1997
Manufacturer
DIAGNOSTIC SOLUTIONS, INC.
Product Code
CFR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED OUT OF RANGE NORMAL CONTROL SOLUTION TEST WITH TEST STRIPS. ON 3/26/97, THE CUSTOMER OPENED THE SECOND BOTTLE FROM A BOX OF 50 AND PERFORMED A NORMAL CONTROL SOLUTION TEST. THE RESULT WAS 19 MG/DL VERSUS A NORMAL CONTROL SOLUTION RANGE OF 122-182 MG/DL. THE CUSTOMER CALLED THE CO CUSTOMER SUPPORT LINE AND, WITH THE REP, PERFORMED A NORMAL CONTROL SOLUTION TEST. THE RESULT WAS 27 MG/DL. A CHECK STRIP TEST WAS PERFORMED WITH THE REP. THE RESULT WAS 84 VERSUS A CHECK STRIP OF 70-93. THE CUSTOMER REMOVED THE DESICCANT UPON OPENING OF THE BOTTLE. THE CUSTOMER STATED THAT HE HAD NOT REMOVED THE DESICCANT FROM THE FIRST BOTTLE AND THAT THE FIRST BOTTLE HAD PERFORMED ACCURATELY. THE CUSTOMER STORES THE TEST STRIPS IN THE DEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK CHECK ONE BLOOD GLUCOSE TEST STRIPS FOR USE WITH ONE TOUCH METERS CFR DIAGNOSTIC SOLUTIONS, INC. * 1A612A

Patients

Seq Age Sex Outcome Treatment
1 *