FDA Adverse Event Malfunction Summary report: N

QUICK CHECK ONE

MDR report key: 86482 · Received April 24, 1997

Report

Report Number
2027242-1997-90134
Event Type
Malfunction
Date Received
April 24, 1997
Date of Event
March 25, 1997
Report Date
March 26, 1997
Manufacturer
DIAGNOSTIC SOLUTIONS, INC.
Product Code
CFR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED OUT OF RANGE NORMAL CONTROL SOLUTION TEST RESULTS WITH TEST STRIPS. ON 3/25/97, THE CUSTOMER OPENED THE 2ND BOTTLE FROM A BOX OF 50 AND PERFORMED A NORMAL CONTROL SOLUTION TEST. THE RESULT WAS 164 MG/DL VERSUS A NORMAL CONTROL SOLUTION RANGE OF 102-154 MG/DL. THE CUSTOMER CALLED THE CO CUSTOMER SUPPORT LINE AND, WITH THE REP, PERFORMED A NORMAL CONTROL SOLUTION TEST. THE RESULT WAS 168 MG/DL. THE CONTROL SOLUTION, LOT # VJ125, EXPIRATION 10/97, WAS OPENED ONE WEEK AGO AND WAS SHAKEN VIGOROUSLY BEFORE THE SAMPLE WAS APPLIED. A CHECK STRIP TEST WAS PERFORMED WITH THE REP. THE RESULT WAS 84 VERSUS A CHECK STRIP RANGE OF 80-106. THE DESICCANT IN IN THE OPEN BOTTLE. THE CUSTOMER STATED THAT THE FIRST BOTTLE PERFORMED ACCURATELY. THE CUSTOMER STORES THE TEST STIPS IN THE BEDROOM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK CHECK ONE BLOOD GLUCOSE TEST STRIPS FOR USE WITH ONE TOUCH METERS CFR DIAGNOSTIC SOLUTIONS, INC. * 1K518B

Patients

Seq Age Sex Outcome Treatment
1 *