UNICEL DXC 800 SYNCHRON CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2007-00010
- Event Type
- Other
- Date Received
- May 2, 2007
- Date of Event
- April 2, 2007
- Report Date
- May 2, 2007
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
PER CUSTOMER, NO QC ISSUES WERE SEEN IN 2007. TWO DAYS LATER, THE CUSTOMER CONDUCTED DAILY MAINTENANCE AND QC, AND OBTAINED LOW FLYERS FOR ONE CONTROL. THE CUSTOMER THEN STARTED RUNNING PT SAMPLES FOR GLU ON A DIFFERENT INSTRUMENT IN THEIR LAB. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S LAB: THE FSE INSPECTED THE INSTRUMENT AND OBSERVED A MALFUNCTIONING STIRRER MOTOR. THE STIRRER MOTOR WAS REPLACED WHICH RESOLVED THE ISSUE. AFTER SERVICE COMPLETION, A PRECISION TESTING WAS PERFORMED WITH ACCEPTABLE RESULTS. A FAILED STIRRER MOTOR ASSEMBLY MAY HAVE CONTRIBUTED TO THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
PER BECKMAN COULTER INC. (BCI) CUSTOMER, INCORRECT GLUCOSE (GLUM) RESULTS WERE REPORTED FOR MULTIPLE PT SAMPLES THAT WERE GENERATED BY THE UNICEL DXC 800 SYNCHRON CLINICAL SYSTEMS. THE CUSTOMER INDICATED THAT ERRONEOUS GLUM RESULTS WERE REPORTED FOR APPROXIMATELY 100 PTS (SEE EXAMPLES IN B6). AFTER SERVICE COMPLETION, THE CUSTOMER RE-TESTED ALL PT SAMPLES FOR GLUM AND AMENDED ABOUT 100 RESULTS (SEE EXAMPLES IN B6). THERE IS NO INFORMATION REGARDING ANY EFFECTS TO PTS, BUT THE CUSTOMER HAS NOT BEEN INFORMED OF ANY INCIDENT RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 SYNCHRON CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |