VITROS 350 CHEMISTRY SYSTEM
Report
- Report Number
- 1319681-2019-00044
- Event Type
- Malfunction
- Date Received
- May 23, 2019
- Date of Event
- March 3, 2019
- Report Date
- May 23, 2019
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT ONE HIGHER AND THREE LOWER THAN EXPECTED AMMONIA RESULTS WERE OBTAINED FROM ORTHO LIQUID PERFORMANCE VERIFIER (LPV) CONTROL FLUIDS WHEN TESTED USING VITROS AMMONIA (AMON) SLIDES ON A VITROS 350 CHEMISTRY SYSTEM. THE INVESTIGATION CONCLUDED THE MOST LIKELY ASSIGNABLE CAUSE IS INSTRUMENT RELATED, ASSOCIATED WITH MICROSLIDE INCUBATOR CONTAMINATION. AFTER CLEANING THE INCUBATOR EVAPORATION CAPS, ACCEPTABLE VITROS AMON PERFORMANCE WAS OBTAINED.
A CUSTOMER OBTAINED ONE HIGHER AND THREE LOWER THAN EXPECTED AMMONIA RESULTS FROM ORTHO LIQUID PERFORMANCE VERIFIER (LPV) CONTROL FLUIDS WHEN TESTED USING VITROS AMMONIA (AMON) SLIDES ON A VITROS 350 CHEMISTRY SYSTEM. ORTHO LPV F6495 RESULT OF 93.4 UMOL/L WAS OBTAINED VERSUS AN EXPECTED RESULT OF 47.1 UMOL/L. ORTHO LPV G6496 RESULTS OF 150.3, 98.5 AND 71.6 UMOL/L WERE OBTAINED VERSUS AN EXPECTED RESULT OF 197.2 UMOL/L. BIASED RESULTS OF THE DIRECTION AND MAGNITUDE OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF UNDETECTED. THERE WAS NO REPORT OF AFFECTED PATIENT RESULTS AND THERE WAS NO ALLEGATION OF ACTUAL PATIENT HARM. HOWEVER, THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS HAD NOT BEEN AFFECTED OR WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO).COMPLAINT NUMBERS (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 432453 | VITROS 350 CHEMISTRY SYSTEM | CHEMISTRY ANALYZER | JJE | ORTHO-CLINICAL DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |