FDA Adverse Event Other Summary report: N

HALL SURGICAL BUR GUARD

MDR report key: 863832 · Received June 5, 2007

Report

Report Number
1017294-2007-00379
Event Type
Other
Date Received
June 5, 2007
Date of Event
April 13, 2007
Report Date
May 15, 2007
Manufacturer
CONMED LINVATEC
Product Code
EQJ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS: THE CUSTOMER RETURNED SEVERAL BUR GUARDS FOR EVALUATION SINCE THE ONE ASSOCIATED WITH A BURN TO THE PATIENT'S LIP COULD NOT BE IDENTIFIED. AN EVALUATION OF THIS BUR GUARD WAS FOUND BEARING DAMAGE. IN ADDITION, RUST-LIKE DEPOSITS WERE NOTED IN THE ASSEMBLY INDICATING POSSIBLE IMMERSION. THE CUSTOMER HAS BEEN PROVIDED ADDITIONAL EDUCATION ON THE USE AND CARE OF THIS PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF A BUR GUARD IN ORAL SURGERY, THE PATIENT RECEIVED A BUR TO THE LIP. THE AFFECTED AREA WAS TREATED WITH ANTIBIOTIC OINTMENT AND REPORTED TO BE HEALING WELL WITHOUT FURTHER TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HALL SURGICAL BUR GUARD BUR GUARD EQJ CONMED LINVATEC NA NOV06

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other| R CAT#: 00602002500 & SERIAL #: UNKNOWN| MICROPOWER ORAL MAX HS DRILL